Dosimetry during iodine-131 therapy - a technical point of view from a single centre's own experience.

BACKGROUND: Nuclear medicine uses radionuclides in medicine for diagnosis, staging, therapy, and monitoring the response to therapy. The application of radiopharmaceutical therapy for the treatment of certain diseases is well-established, and the field is expanding. Internal dosimetry is multifaceted and includes different workflows, as well as various calculations based on patient- specific dosimetry.

AIM: The objective of this study was to introduce the technical issues which might occur during iodine-131 (¹³¹I) dosimetry performed in nuclear medicine departments.

MATERIAL AND METHODS: Retrospective analysis was performed on a group of 44 patients with papillary thyroid cancer who between May 2021 and October 2021 underwent a 131I treatment: 80-100 mCi (2200-3700 MBq, based on the previous medical history and stage of the disease). Patients underwent a series of ¹³¹I therapy scans using gamma camera Discovery NM 670 CT. Whole body scan (WBS) was performed 2, 4, 24 and 48 hours after ¹³¹I administration. Additionally, after 24 hours of single photon emission computed tomography/ computed tomography, two fields of view (SPECT/CT 2-FOV) were performed from the mid-head to the bladder.

RESULTS: During the dosimetry procedure, several issues arise. Firstly, after receiving therapeutic doses of ¹³¹I, patients should remain in their rooms until the appropriate activity is achieved before being transported to the diagnostic room. Secondly, the walls between examination rooms meet the requirements for accurate diagnosis but not for therapy, leading to the occurrence of artefacts in patients examined behind the wall, potentially influencing the examination results. Thirdly, personnel in the control room also experience additional exposure (10 times greater than in the case of standard diagnostic procedure).

CONCLUSIONS: The dosimetry in patients in whom therapeutic procedures are performed with the use of isotopes is mandatory according to Polish and European law, technical issues which occur during the dosimetry procedures might influence the organization of the work in departments.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app