Intrasession repeatability and agreement of the anterior corneal assessment provided by a multidiagnostic device.
Clinical & Experimental Optometry : Journal of the Australian Optometrical Association 2024 March 18
CLINICAL RELEVANCE: Multidiagnostic systems have recently appeared on the market. Knowledge of the repeatability and validity of any instrument is mandatory before its introduction in clinical practice.
BACKGROUND: The aim of this work is to examine the intrasession repeatability of anterior pole measurements provided by the multidiagnostic device Wave Analyzer Medica 700 (WAM700) and agreement with Pentacam measurements in normal eyes.
METHODS: In the right eyes of 113 participants, three repeat measurements of central keratometry, central corneal thickness, anterior chamber depth and corneal eccentricity were made with the WAM700 and Pentacam in random order. Intrasession repeatability and agreement were determined.
RESULTS: Employing WAM700, intrasession repeatability for keratometry, central corneal thickness and anterior chamber depth was good (ICCs ≥ 0.992; CV 0.48-0.98%), yet worse than the values obtained for the Pentacam (ICCs ≥ 0.998; CV 0-0.33%). WAM700 showed excellent intrasession repeatability when used to measure the anterior chamber depth (Sw 0.03 mm). However, the repeatability of this device was inferior for central corneal thickness (Sw 4.24 μm) and keratometry measurements (Sw < 0.21 D) and was poor for corneal eccentricity (Sw 0.07; ICC 0.908; CV 14.58%). Agreement between WAM700 and Pentacam showed a high ICC for the keratometry measurements, central corneal thickness and anterior chamber depth (>0.972) but lower for corneal eccentricity (ICC 0.762).
CONCLUSIONS: In healthy eyes, the WAM700 multidiagnostic device showed good intrasession repeatability for keratometry, central corneal thickness and anterior chamber depth measurements. Agreement between WAM700 and Pentacam was good for the anterior chamber depth measurement. However, these instruments cannot be considered interchangeable for keratometry, central corneal thickness and eccentricity readings.
BACKGROUND: The aim of this work is to examine the intrasession repeatability of anterior pole measurements provided by the multidiagnostic device Wave Analyzer Medica 700 (WAM700) and agreement with Pentacam measurements in normal eyes.
METHODS: In the right eyes of 113 participants, three repeat measurements of central keratometry, central corneal thickness, anterior chamber depth and corneal eccentricity were made with the WAM700 and Pentacam in random order. Intrasession repeatability and agreement were determined.
RESULTS: Employing WAM700, intrasession repeatability for keratometry, central corneal thickness and anterior chamber depth was good (ICCs ≥ 0.992; CV 0.48-0.98%), yet worse than the values obtained for the Pentacam (ICCs ≥ 0.998; CV 0-0.33%). WAM700 showed excellent intrasession repeatability when used to measure the anterior chamber depth (Sw 0.03 mm). However, the repeatability of this device was inferior for central corneal thickness (Sw 4.24 μm) and keratometry measurements (Sw < 0.21 D) and was poor for corneal eccentricity (Sw 0.07; ICC 0.908; CV 14.58%). Agreement between WAM700 and Pentacam showed a high ICC for the keratometry measurements, central corneal thickness and anterior chamber depth (>0.972) but lower for corneal eccentricity (ICC 0.762).
CONCLUSIONS: In healthy eyes, the WAM700 multidiagnostic device showed good intrasession repeatability for keratometry, central corneal thickness and anterior chamber depth measurements. Agreement between WAM700 and Pentacam was good for the anterior chamber depth measurement. However, these instruments cannot be considered interchangeable for keratometry, central corneal thickness and eccentricity readings.
Full text links
Related Resources
Trending Papers
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Prevention and treatment of ischaemic and haemorrhagic stroke in people with diabetes mellitus: a focus on glucose control and comorbidities.Diabetologia 2024 April 17
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Eosinophilic Esophagitis: Clinical Pearls for Primary Care Providers and Gastroenterologists.Mayo Clinic Proceedings 2024 April
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app