Embolization of percutaneous left atrial appendage closure devices: timing, management and clinical outcomes.

BACKGROUND: Left atrial appendage (LAA) occluder embolization is an infrequent but serious complication.

OBJECTIVES: We aim to describe timing, management and clinical outcomes of device embolization in a multi-center registry.

METHODS: Patient characteristics, imaging findings and procedure and follow-up data were collected retrospectively. Device embolizations were categorized according to 1) timing 2) management and 3) clinical outcomes.

RESULTS: Sixty-seven centers contributed data. Device embolization occurred in 108 patients. In 70.4 % of cases, it happened within the first 24 h of the procedure. The device was purposefully left in the LA and the aorta in two (1.9 %) patients, an initial percutaneous retrieval was attempted in 81 (75.0 %) and surgery without prior percutaneous retrieval attempt was performed in 23 (21.3 %) patients. Two patients died before a retrieval attempt could be made. In 28/81 (34.6 %) patients with an initial percutaneous retrieval attempt a second, additional attempt was performed, which was associated with a high mortality (death in patients with one attempt: 2.9 % vs. second attempt: 21.4 %, p < 0.001). The primary outcome (bailout surgery, cardiogenic shock, stroke, TIA, and/or death) occurred in 47 (43.5 %) patients. Other major complications related to device embolization occurred in 21 (19.4 %) patients.

CONCLUSIONS: The majority of device embolizations after LAA closure occurs early. A percutaneous approach is often the preferred method for a first rescue attempt. Major adverse event rates, including death, are high particularly if the first retrieval attempt was unsuccessful.

CONDENSED ABSTRACT: This dedicated multicenter registry examined timing, management, and clinical outcome of device embolization. Early embolization (70.4 %) was most frequent. As a first rescue attempt, percutaneous retrieval was preferred in 75.0 %, followed by surgical removal (21.3 %). In patients with a second retrieval attempt a higher mortality (death first attempt: 2.9 % vs. death second attempt: 24.1 %, p < 0.001) was observed. Mortality (10.2 %) and the major complication rate after device embolization were high.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app