An implantable magnetic drive mechanism for non-invasive arteriovenous conduit blood flow control.
IEEE Transactions on Bio-medical Engineering 2024 Februrary 28
OBJECTIVE: Hemodialysis patients usually receive an arteriovenous fistula (AVF) in the arm as vascular access conduit to allow dialysis 2-3 times a week. This AVF introduces the high flow necessary for dialysis, but over time the ever-present supraphysiological flow is the leading cause of complications. This study aims to develop an implantable device able to non-invasively remove the high flow outside dialysis sessions.
METHODS: The developed prototype features a magnetic ring allowing external coupling and torque transmission to non-invasively control an AVF valve. Mock-up devices were implanted into arm and sheep cadavers to test sizes and locations. The transmission torque, output force, and valve closure are measured for different representative skin thicknesses.
RESULTS: The prototype was placed successfully into arm and sheep cadavers. In the prototype, a maximum output force of 78.9±4.2 N, 46.7±1.9 N, 25.6±0.7 N, 13.5±0.6 N and 6.3±0.4 N could be achieved non-invasively through skin thicknesses of 1-5 mm respectively. The fistula was fully collapsible in every measurement through skin thickness up to the required 4 mm.
CONCLUSION: The prototype satisfies the design requirements. It is fully implantable and allows closure and control of an AVF through non-invasive torque transmission. In vivo studies are pivotal in assessing functionality and understanding systemic effects.
SIGNIFICANCE: A method is introduced to transfer large amounts of energy to a medical implant for actuation of a mechanical valve trough a closed surface. This system allows non-invasive control of an AVF to reduce complications related to the permanent high flow in conventional AVFs.
METHODS: The developed prototype features a magnetic ring allowing external coupling and torque transmission to non-invasively control an AVF valve. Mock-up devices were implanted into arm and sheep cadavers to test sizes and locations. The transmission torque, output force, and valve closure are measured for different representative skin thicknesses.
RESULTS: The prototype was placed successfully into arm and sheep cadavers. In the prototype, a maximum output force of 78.9±4.2 N, 46.7±1.9 N, 25.6±0.7 N, 13.5±0.6 N and 6.3±0.4 N could be achieved non-invasively through skin thicknesses of 1-5 mm respectively. The fistula was fully collapsible in every measurement through skin thickness up to the required 4 mm.
CONCLUSION: The prototype satisfies the design requirements. It is fully implantable and allows closure and control of an AVF through non-invasive torque transmission. In vivo studies are pivotal in assessing functionality and understanding systemic effects.
SIGNIFICANCE: A method is introduced to transfer large amounts of energy to a medical implant for actuation of a mechanical valve trough a closed surface. This system allows non-invasive control of an AVF to reduce complications related to the permanent high flow in conventional AVFs.
Full text links
Related Resources
Trending Papers
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Prevention and treatment of ischaemic and haemorrhagic stroke in people with diabetes mellitus: a focus on glucose control and comorbidities.Diabetologia 2024 April 17
British Society for Rheumatology guideline on management of adult and juvenile onset Sjögren disease.Rheumatology 2024 April 17
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Albumin: a comprehensive review and practical guideline for clinical use.European Journal of Clinical Pharmacology 2024 April 13
Eosinophilic Esophagitis: Clinical Pearls for Primary Care Providers and Gastroenterologists.Mayo Clinic Proceedings 2024 April
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app