We have located links that may give you full text access.
Accuracy of Two Continuous Glucose Monitoring Devices During Aerobic and High-Intensity Interval Training in Individuals with Type 1 Diabetes.
Diabetes Technology & Therapeutics 2024 January 13
BBACKGROUND: This study aimed to evaluate the accuracy of Dexcom G6 (DG6) and FreeStyle Libre-2 (FSL2) during aerobic training and HIIT in individuals with type 1 diabetes (T1D).
METHODS: Twenty-six males (mean age 29.3 ± 6.3 years and mean duration of diabetes 14.9 ± 6.1 years) participated in this study. Interstitial glucose levels were measured using DG6 and FSL2, while plasma glucose levels were measured every 10 min using YSI 2500 as the reference for glucose measurements in this study. The measurements began 20 min before the start of exercise and continued for 20 min after exercise. Seven measurements were taken for each subject and exercise.
RESULTS: Both DG6 and FSL2 devices showed significant differences compared to YSI glucose data for both aerobic and HIIT exercises. Continuous glucose monitoring (CGM) devices exhibited superior performance during HIIT than aerobic training, with DG6 showing a mean absolute relative difference (MARD) of 14.03% versus 31.98%, respectively. In the comparison between the two devices, FSL2 demonstrated significantly higher effectiveness in aerobic training, yet its performance was inferior to DG6 during HIIT. According to the 40/40 criteria, both sensors performed similarly, with marks over 93% for all ranges and both exercises, and above 99% for HIIT and in the >180 mg/dL range, which is in accordance with FDA guidelines.
CONCLUSIONS: The findings suggest that the accuracy of DG6 and FSL2 deteriorates during and immediately after exercise, but remains acceptable for both devices during HIIT. However, accuracy is compromised with DG6 during aerobic exercise. This study is the first to compare the accuracy of two CGMs, DG6 and FSL2, during two exercise modalities, using plasma glucose YSI measurements as the gold standard for comparisons.
METHODS: Twenty-six males (mean age 29.3 ± 6.3 years and mean duration of diabetes 14.9 ± 6.1 years) participated in this study. Interstitial glucose levels were measured using DG6 and FSL2, while plasma glucose levels were measured every 10 min using YSI 2500 as the reference for glucose measurements in this study. The measurements began 20 min before the start of exercise and continued for 20 min after exercise. Seven measurements were taken for each subject and exercise.
RESULTS: Both DG6 and FSL2 devices showed significant differences compared to YSI glucose data for both aerobic and HIIT exercises. Continuous glucose monitoring (CGM) devices exhibited superior performance during HIIT than aerobic training, with DG6 showing a mean absolute relative difference (MARD) of 14.03% versus 31.98%, respectively. In the comparison between the two devices, FSL2 demonstrated significantly higher effectiveness in aerobic training, yet its performance was inferior to DG6 during HIIT. According to the 40/40 criteria, both sensors performed similarly, with marks over 93% for all ranges and both exercises, and above 99% for HIIT and in the >180 mg/dL range, which is in accordance with FDA guidelines.
CONCLUSIONS: The findings suggest that the accuracy of DG6 and FSL2 deteriorates during and immediately after exercise, but remains acceptable for both devices during HIIT. However, accuracy is compromised with DG6 during aerobic exercise. This study is the first to compare the accuracy of two CGMs, DG6 and FSL2, during two exercise modalities, using plasma glucose YSI measurements as the gold standard for comparisons.
Full text links
Related Resources
Trending Papers
Consensus Statement on Vitamin D Status Assessment and Supplementation: Whys, Whens, and Hows.Endocrine Reviews 2024 April 28
The Tricuspid Valve: A Review of Pathology, Imaging, and Current Treatment Options: A Scientific Statement From the American Heart Association.Circulation 2024 April 26
Intravenous infusion of dexmedetomidine during the surgery to prevent postoperative delirium and postoperative cognitive dysfunction undergoing non-cardiac surgery: a meta-analysis of randomized controlled trials.European Journal of Medical Research 2024 April 19
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Ventilator Waveforms May Give Clues to Expiratory Muscle Activity.American Journal of Respiratory and Critical Care Medicine 2024 April 25
Acute Kidney Injury and Electrolyte Imbalances Caused by Dapagliflozin Short-Term Use.Pharmaceuticals 2024 March 27
Systemic lupus erythematosus.Lancet 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app