Clinical validation of an AI-based automatic quantification tool for lung lobes in SPECT/CT.

EJNMMI Physics 2023 September 22
BACKGROUND: Lung lobar ventilation and perfusion (V/Q) quantification is generally obtained by generating planar scintigraphy images and then imposing three equally sized regions of interest on the data of each lung. This method is fast but not as accurate as SPECT/CT imaging, which provides three-dimensional data and therefore allows more precise lobar quantification. However, the manual delineation of each lobe is time-consuming, which makes SPECT/CT incompatible with the clinical workflow for V/Q estimation. An alternative may be to use artificial intelligence-based auto-segmentation tools such as AutoLung3D (Siemens Healthineers, Knoxville, USA), which automatically delineate the lung lobes on the CT data acquired with the SPECT data. The present study assessed the clinical validity of this approach relative to planar scintigraphy and manual quantification in SPECT/CT.

METHODS: The Autolung3D software was tested on the retrospective SPECT/CT data of 43 patients who underwent V/Q scintigraphy with 99m Tc-macroaggregated albumin and 99m Tc-labeled aerosol. It was compared to planar scintigraphy and SPECT/CT using the manual quantification method in terms of relative lobar V/Q quantification values and interobserver variability.

RESULTS: The three methods provided similar V/Q estimates for the left lung lobes and total lungs. However, compared to the manual SPECT/CT method, planar scintigraphy yielded significantly higher estimates for the middle right lobe and significantly lower estimates for the superior and inferior right lobes. The estimates of the manual and automated SPECT/CT methods were similar. However, the post-processing time in the automated method was approximately 5 min compared to 2 h for the manual method. Moreover, the automated method associated with a drastic reduction in interobserver variability: Its maximal relative standard deviation was only 5%, compared to 23% for planar scintigraphy and 19% for the manual SPECT/CT method.

CONCLUSIONS: This study validated the AutoLung3D software for general clinical use since it rapidly provides accurate lobar quantification in V/Q scans with markedly less interobserver variability than planar scintigraphy or the manual SPECT/CT method.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app