We have located links that may give you full text access.
Clinical evaluation the success rate and complications of fluoroscopically guided removal of tracheal tube metallic stents.
Journal of Cardiothoracic Surgery 2021 March 26
BACKGROUND: Long-term placement of airway stents has a high probability of restenosis of the airway due to granulation tissue hyperplasia, and it is difficult to remove the stent. Our aim is to evaluate the success rate and complications of removal of tracheal tube metallic stents under fluoroscopic guidance, and to compare the difference between uncovered stent and covered stent.
METHODS: We retrospectively reviewed 45 cases (31 males and 14 females; age, 12-71 years) of tracheal metallic stent removal performed at our center between January 2014 and December 2019. Covered stents were applied in 36 cases, and uncovered stents were applied in 9 cases. In the covered stent group, 15 patients presented with granulation tissue at both ends; 3 cases, with stent fracture; and 2, with stent intolerance due to severe airway foreign body sensation. In the uncovered stents group, all patients presented with granulation tissue formation; 2 patients, with stent fracture; and 1 patient, with stent intolerance.
RESULTS: A total of 41 (91.1%) stents were successfully removed (34 [94.4%] in the covered stent group and 7 [77.8%] in the uncovered stent group). The average duration of stent placement was 3.2 ± 0.7 and 2.5 ± 1.2 months in the covered stent group and uncovered stent group, respectively. With regard to the complications, hemoptysis occurred in 4 cases (average blood volume lost, 100 ml), tracheal mucosa tear occurred in 5 cases, tracheal collapse requiring emergency airway stent placement occurred in 1 case, and tracheal rupture requiring emergency surgical suture occurred in 1 case. No procedure-related deaths occurred in either group.
CONCLUSIONS: It is safe to remove the metal stent of the tracheal tube under the guidance of fluoroscopy, with low complications, and can avoid the long-term placement of the airway stent.
METHODS: We retrospectively reviewed 45 cases (31 males and 14 females; age, 12-71 years) of tracheal metallic stent removal performed at our center between January 2014 and December 2019. Covered stents were applied in 36 cases, and uncovered stents were applied in 9 cases. In the covered stent group, 15 patients presented with granulation tissue at both ends; 3 cases, with stent fracture; and 2, with stent intolerance due to severe airway foreign body sensation. In the uncovered stents group, all patients presented with granulation tissue formation; 2 patients, with stent fracture; and 1 patient, with stent intolerance.
RESULTS: A total of 41 (91.1%) stents were successfully removed (34 [94.4%] in the covered stent group and 7 [77.8%] in the uncovered stent group). The average duration of stent placement was 3.2 ± 0.7 and 2.5 ± 1.2 months in the covered stent group and uncovered stent group, respectively. With regard to the complications, hemoptysis occurred in 4 cases (average blood volume lost, 100 ml), tracheal mucosa tear occurred in 5 cases, tracheal collapse requiring emergency airway stent placement occurred in 1 case, and tracheal rupture requiring emergency surgical suture occurred in 1 case. No procedure-related deaths occurred in either group.
CONCLUSIONS: It is safe to remove the metal stent of the tracheal tube under the guidance of fluoroscopy, with low complications, and can avoid the long-term placement of the airway stent.
Full text links
Related Resources
Trending Papers
Consensus Statement on Vitamin D Status Assessment and Supplementation: Whys, Whens, and Hows.Endocrine Reviews 2024 April 28
The Tricuspid Valve: A Review of Pathology, Imaging, and Current Treatment Options: A Scientific Statement From the American Heart Association.Circulation 2024 April 26
Intravenous infusion of dexmedetomidine during the surgery to prevent postoperative delirium and postoperative cognitive dysfunction undergoing non-cardiac surgery: a meta-analysis of randomized controlled trials.European Journal of Medical Research 2024 April 19
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Ventilator Waveforms May Give Clues to Expiratory Muscle Activity.American Journal of Respiratory and Critical Care Medicine 2024 April 25
Acute Kidney Injury and Electrolyte Imbalances Caused by Dapagliflozin Short-Term Use.Pharmaceuticals 2024 March 27
Systemic lupus erythematosus.Lancet 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app