Comparison of Barricor™ vs. lithium heparin tubes for selected routine biochemical analytes and evaluation of post centrifugation stability.
Introduction: Obtaining suitable results unaffected by pre- or postanalytical phases is pivotal for clinical chemistry service. We aimed comparison and stability of nine biochemical analytes after centrifugation using Barricor™ plasma tubes with mechanical separator vs standard Vacutainer® lithium heparin tubes.
Materials and methods: We collected samples on six healthy volunteers and nine patients from intensive care units into 6 mL plastic Vacutainer® lithium heparin tubes and 5.5 mL plastic Barricor™ plasma tubes. All tubes were centrifuged within 30 minutes after venipuncture. First, we compared results of nine biochemical analytes from lithium heparin tubes with Barricor™ tubes for each analyte using Passing-Bablok and Bland-Altman analyses. Second, we calculated the difference of analyte concentrations between baseline and time intervals in tubes stored at + 4 °C. Based on the total change limit we calculated the maximum allowable concentrations percentage change from baseline.
Results: The majority of correlation coefficients were close to 0.99 indicating good correlation in the working range. Bland-Altman analyses showed an acceptable concordance for all analytes. In consequence, the Barricor™ tube might be an alternative to regular lithium heparin tube. Stability with this new generation tube is improved for eight analytes (except for aspartate aminotransferase) in comparison with regular lithium heparin tubes.
Conclusions: By using Barricor™ tubes and prompt centrifugation, supplemental analysis or re-analysis for eight analytes including alanine aminotransferase, alkaline phosphatase, C-reactive protein, high sensitivity troponin T, lactate dehydrogenase, NT-pro BNP, potassium and sodium could be performed within 72 h of specimen collection.
Materials and methods: We collected samples on six healthy volunteers and nine patients from intensive care units into 6 mL plastic Vacutainer® lithium heparin tubes and 5.5 mL plastic Barricor™ plasma tubes. All tubes were centrifuged within 30 minutes after venipuncture. First, we compared results of nine biochemical analytes from lithium heparin tubes with Barricor™ tubes for each analyte using Passing-Bablok and Bland-Altman analyses. Second, we calculated the difference of analyte concentrations between baseline and time intervals in tubes stored at + 4 °C. Based on the total change limit we calculated the maximum allowable concentrations percentage change from baseline.
Results: The majority of correlation coefficients were close to 0.99 indicating good correlation in the working range. Bland-Altman analyses showed an acceptable concordance for all analytes. In consequence, the Barricor™ tube might be an alternative to regular lithium heparin tube. Stability with this new generation tube is improved for eight analytes (except for aspartate aminotransferase) in comparison with regular lithium heparin tubes.
Conclusions: By using Barricor™ tubes and prompt centrifugation, supplemental analysis or re-analysis for eight analytes including alanine aminotransferase, alkaline phosphatase, C-reactive protein, high sensitivity troponin T, lactate dehydrogenase, NT-pro BNP, potassium and sodium could be performed within 72 h of specimen collection.
Full text links
Related Resources
Trending Papers
British Society of Gastroenterology guidelines for the management of hepatocellular carcinoma in adults.Gut 2024 April 17
Systemic lupus erythematosus.Lancet 2024 April 18
Should renin-angiotensin system inhibitors be held prior to major surgery?British Journal of Anaesthesia 2024 May
Ventilator Waveforms May Give Clues to Expiratory Muscle Activity.American Journal of Respiratory and Critical Care Medicine 2024 April 25
Acute Kidney Injury and Electrolyte Imbalances Caused by Dapagliflozin Short-Term Use.Pharmaceuticals 2024 March 27
Colorectal polypectomy and endoscopic mucosal resection: European Society of Gastrointestinal Endoscopy (ESGE) Guideline - Update 2024.Endoscopy 2024 April 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app