Evaluation of a newly developed chemiluminescence immunoassay for detecting cardiac troponin T.
Journal of Clinical Laboratory Analysis 2018 March
BACKGROUND: To evaluate the performance of a chemiluminescence detection kit for cardiac troponin T (cTnT).
METHODS: According to the "Guiding principles on performance analysis of diagnostic reagents in vitro" and the Clinical and Laboratory Standards Institute (CLSI) Guidelines, we assessed the detection limit, linear range, reportable range, accuracy, precision, cross reactivity, interference factors, and matrix effect of the kit, and compared these parameters with that of the commercial electrochemiluminescence detection kit for cTnT (Roche).
RESULTS: The minimum detection limit of the kit was 0.01 ng/mL. The linear range was 0.01 ng/mL-25 ng/mL. The reportable range was from 0.01 ng/mL to 100 ng/mL. Precision within the batch was 2.9%-6.4%, and precision between batches was 6.0%; the accuracy was good and the recovery rate reached 96.2%. The cross-reaction test demonstrated that there was no reactivity between cTnT and a variety of troponins. Test results deviated by less than ±10% in the presence of hemoglobin ≤1000 μg/mL, bilirubin ≤250 μg/mL, triglycerides ≤11.25 mg/mL, and rheumatoid factor ≤206 U/mL, suggesting that kit results were not significantly interfered with these factors. Results from the matrix-effect assessment revealed absence of a matrix effect in the tested serum samples. Correlation study revealed that the performance of the kit was very consistent with that of the Roche electrochemiluminescence detection kit (Kappa = 0.900, P < .001) with a high correlation (r = .903, P < .001) and a total matching rate of 95.00%.
CONCLUSION: The performance of the evaluated chemiluminescence immunoquantitation kit for cTnT detection was acceptable in all tested parameters, fulfilling requirements for clinical applications.
METHODS: According to the "Guiding principles on performance analysis of diagnostic reagents in vitro" and the Clinical and Laboratory Standards Institute (CLSI) Guidelines, we assessed the detection limit, linear range, reportable range, accuracy, precision, cross reactivity, interference factors, and matrix effect of the kit, and compared these parameters with that of the commercial electrochemiluminescence detection kit for cTnT (Roche).
RESULTS: The minimum detection limit of the kit was 0.01 ng/mL. The linear range was 0.01 ng/mL-25 ng/mL. The reportable range was from 0.01 ng/mL to 100 ng/mL. Precision within the batch was 2.9%-6.4%, and precision between batches was 6.0%; the accuracy was good and the recovery rate reached 96.2%. The cross-reaction test demonstrated that there was no reactivity between cTnT and a variety of troponins. Test results deviated by less than ±10% in the presence of hemoglobin ≤1000 μg/mL, bilirubin ≤250 μg/mL, triglycerides ≤11.25 mg/mL, and rheumatoid factor ≤206 U/mL, suggesting that kit results were not significantly interfered with these factors. Results from the matrix-effect assessment revealed absence of a matrix effect in the tested serum samples. Correlation study revealed that the performance of the kit was very consistent with that of the Roche electrochemiluminescence detection kit (Kappa = 0.900, P < .001) with a high correlation (r = .903, P < .001) and a total matching rate of 95.00%.
CONCLUSION: The performance of the evaluated chemiluminescence immunoquantitation kit for cTnT detection was acceptable in all tested parameters, fulfilling requirements for clinical applications.
Full text links
Related Resources
Trending Papers
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Prevention and treatment of ischaemic and haemorrhagic stroke in people with diabetes mellitus: a focus on glucose control and comorbidities.Diabetologia 2024 April 17
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Eosinophilic Esophagitis: Clinical Pearls for Primary Care Providers and Gastroenterologists.Mayo Clinic Proceedings 2024 April
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app