Allergic contact dermatitis caused by isobornyl acrylate in Freestyle® Libre, a newly introduced glucose sensor.
Contact Dermatitis 2017 December
BACKGROUND: Glucose sensors, such as FreeStyle® Libre, are innovative medical devices developed for diabetes patients as a replacement for classic glucose meters, ensuring continuous glucose monitoring without the disadvantage of regular skin finger pricks.
OBJECTIVES: To report several cases of allergic contact dermatitis caused by FreeStyle® Libre, and to report on isobornyl acrylate as a culprit allergen.
PATIENTS AND METHODS: Fifteen patients presented with allergic contact dermatitis caused by FreeStyle® Libre. All but 1 were patch tested with a baseline series, and with pieces and/or ultrasonic bath extracts of (the adhesive part of) the glucose sensor. Isobornyl acrylate was patch tested, in various concentrations and vehicles, in 13 patients. Gas chromatography-mass spectrometry (GC-MS) of the sensors was performed.
RESULTS: All patients reacted to the adhesive part of the sensor, and 12 patients were shown to be sensitized to isobornyl acrylate. Simultaneous reactions to other allergens were rarely observed. GC-MS showed the presence of isobornyl acrylate in the sensors.
CONCLUSIONS: Cases of allergic contact dermatitis caused by FreeStyle® Libre are increasingly being observed, and isobornyl acrylate is a relevant culprit allergen. Cross-reactivity to other acrylates was infrequently observed, but other, hitherto unidentified, contact allergens may still be present in the device.
OBJECTIVES: To report several cases of allergic contact dermatitis caused by FreeStyle® Libre, and to report on isobornyl acrylate as a culprit allergen.
PATIENTS AND METHODS: Fifteen patients presented with allergic contact dermatitis caused by FreeStyle® Libre. All but 1 were patch tested with a baseline series, and with pieces and/or ultrasonic bath extracts of (the adhesive part of) the glucose sensor. Isobornyl acrylate was patch tested, in various concentrations and vehicles, in 13 patients. Gas chromatography-mass spectrometry (GC-MS) of the sensors was performed.
RESULTS: All patients reacted to the adhesive part of the sensor, and 12 patients were shown to be sensitized to isobornyl acrylate. Simultaneous reactions to other allergens were rarely observed. GC-MS showed the presence of isobornyl acrylate in the sensors.
CONCLUSIONS: Cases of allergic contact dermatitis caused by FreeStyle® Libre are increasingly being observed, and isobornyl acrylate is a relevant culprit allergen. Cross-reactivity to other acrylates was infrequently observed, but other, hitherto unidentified, contact allergens may still be present in the device.
Full text links
Related Resources
Trending Papers
British Society of Gastroenterology guidelines for the management of hepatocellular carcinoma in adults.Gut 2024 April 17
Systemic lupus erythematosus.Lancet 2024 April 18
Should renin-angiotensin system inhibitors be held prior to major surgery?British Journal of Anaesthesia 2024 May
Ventilator Waveforms May Give Clues to Expiratory Muscle Activity.American Journal of Respiratory and Critical Care Medicine 2024 April 25
Acute Kidney Injury and Electrolyte Imbalances Caused by Dapagliflozin Short-Term Use.Pharmaceuticals 2024 March 27
Colorectal polypectomy and endoscopic mucosal resection: European Society of Gastrointestinal Endoscopy (ESGE) Guideline - Update 2024.Endoscopy 2024 April 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app