Comparative Study
Journal Article
Comparison of Spectra Optia and COBE Spectra apheresis systems' performances for red blood cell exchange procedures.
Transfusion and Apheresis Science 2016 December
BACKGROUND: Spectra Optia (Terumo BCT, Lakewood, CO) was FDA approved for red blood cell exchange (RBCx) procedures in January 2014 and is expected to replace COBE spectra (Terumo BCT) very soon in the USA. The performance characteristics of these devices for Isovolemic Hemodilution (IHD-RBCx) procedure were compared in this study.
METHODS: A total of 114 IHD-RBCx procedures from 19 patients were analyzed. For every patient, three procedures on each device with similar pre-procedure hematocrits were compared. Pre and post procedure laboratory parameters compared were hemoglobin S (HbS), hematocrits (Hct), platelet counts and fraction of cells remaining (FCR). Statistical analysis was performed using t-test adjusted by the Holm-Bonferroni method to reduce family-wise error rate.
RESULTS: There were no significant differences between these two devices in regards to HbS, Hct, FCR and platelet counts (p = > 0.05). However, rinseback volume (124.2 ± 8.9 ml) and normal saline replacement volume during IHD phase (296.1 ± 97.2 ml) were lower in Spectra Optia as compared to COBE Spectra (337 ± 33.8 ml and 326.6 ± 105.2 ml, p value <0.001 and 0.030 respectively). Spectra Optia had a longer run time (107.1 ± 15.9 min vs 123.8 ± 19.6 min, p value <0.001) overall.
CONCLUSIONS: Performance characteristics of Spectra Optia for HbS, Hct and FCR were similar to COBE Spectra for IHD-RBCx. IHD-RBCx procedure on Optia required less normal saline replacement volume and rinse back volume but with overall longer procedure run time.
METHODS: A total of 114 IHD-RBCx procedures from 19 patients were analyzed. For every patient, three procedures on each device with similar pre-procedure hematocrits were compared. Pre and post procedure laboratory parameters compared were hemoglobin S (HbS), hematocrits (Hct), platelet counts and fraction of cells remaining (FCR). Statistical analysis was performed using t-test adjusted by the Holm-Bonferroni method to reduce family-wise error rate.
RESULTS: There were no significant differences between these two devices in regards to HbS, Hct, FCR and platelet counts (p = > 0.05). However, rinseback volume (124.2 ± 8.9 ml) and normal saline replacement volume during IHD phase (296.1 ± 97.2 ml) were lower in Spectra Optia as compared to COBE Spectra (337 ± 33.8 ml and 326.6 ± 105.2 ml, p value <0.001 and 0.030 respectively). Spectra Optia had a longer run time (107.1 ± 15.9 min vs 123.8 ± 19.6 min, p value <0.001) overall.
CONCLUSIONS: Performance characteristics of Spectra Optia for HbS, Hct and FCR were similar to COBE Spectra for IHD-RBCx. IHD-RBCx procedure on Optia required less normal saline replacement volume and rinse back volume but with overall longer procedure run time.
Full text links
Related Resources
Trending Papers
Consensus Statement on Vitamin D Status Assessment and Supplementation: Whys, Whens, and Hows.Endocrine Reviews 2024 April 28
The Tricuspid Valve: A Review of Pathology, Imaging, and Current Treatment Options: A Scientific Statement From the American Heart Association.Circulation 2024 April 26
Intravenous infusion of dexmedetomidine during the surgery to prevent postoperative delirium and postoperative cognitive dysfunction undergoing non-cardiac surgery: a meta-analysis of randomized controlled trials.European Journal of Medical Research 2024 April 19
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Ventilator Waveforms May Give Clues to Expiratory Muscle Activity.American Journal of Respiratory and Critical Care Medicine 2024 April 25
Acute Kidney Injury and Electrolyte Imbalances Caused by Dapagliflozin Short-Term Use.Pharmaceuticals 2024 March 27
Systemic lupus erythematosus.Lancet 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app