Validation data for the determination of perchlorate in water using ion chromatography with suppressed conductivity detection.
BACKGROUND: Perchlorate salts are relatively stable, soluble in water, and migrate into groundwater sources. Groundwater is an essential source for drinking water suppliers. Perchlorate bears health risks as it is identified to impair normal thyroid function by interfering with iodine uptake by the thyroid gland. The development of a sensitive analytical method for the determination of perchlorate is therefore of the highest interest or public health. Ion chromatography is a sensitive method suitable for perchlorate determinations. This manuscript describes the validation of an ion chromatographic method. Perchlorate is determined by ion chromatography (IC) with conductivity detection after suppression (CD) applying isocratic elution.
RESULTS: In this study, the suitability of IC-CD was tested for synthetic samples, selected environmental water, drinking water, and swimming pool water in order to evaluate potential matrix effects on the perchlorate signal even after sample preparation. A sample injection volume of 750 μL was applied to the selected 2-mm-IC column. In untreated samples, the perchlorate peak can be interfered by neighbouring signals from matrix ions like chloride, nitrate, carbonate, and sulphate. Depending on the concentration of the matrix ions, the perchlorate peak can show asymmetric shape in particular when the perchlorate concentration is low. Recovery is reduced with increasing matrix ion concentrations. Dedicated matrix elimination was applied to minimize such effects. A reporting limit of 1.5 μg/L perchlorate and an expanded measurement uncertainty of 13.2 % were achieved.
CONCLUSION: The extended method validation proves the applicability of IC based on the EPA 314.0 method for the determination of trace amounts of perchlorate in water samples of different origin. The results support the development of a respective international standard pursued by ISO. The approach evidenced its working robustness and ease of use in terms of eluent preparation, chromatographic resolution, column life time and sample preparation. Due to the simplified analytical workflow of the analytical procedure the application's integration into the collection of methods of interested laboratories should be facilitated.
RESULTS: In this study, the suitability of IC-CD was tested for synthetic samples, selected environmental water, drinking water, and swimming pool water in order to evaluate potential matrix effects on the perchlorate signal even after sample preparation. A sample injection volume of 750 μL was applied to the selected 2-mm-IC column. In untreated samples, the perchlorate peak can be interfered by neighbouring signals from matrix ions like chloride, nitrate, carbonate, and sulphate. Depending on the concentration of the matrix ions, the perchlorate peak can show asymmetric shape in particular when the perchlorate concentration is low. Recovery is reduced with increasing matrix ion concentrations. Dedicated matrix elimination was applied to minimize such effects. A reporting limit of 1.5 μg/L perchlorate and an expanded measurement uncertainty of 13.2 % were achieved.
CONCLUSION: The extended method validation proves the applicability of IC based on the EPA 314.0 method for the determination of trace amounts of perchlorate in water samples of different origin. The results support the development of a respective international standard pursued by ISO. The approach evidenced its working robustness and ease of use in terms of eluent preparation, chromatographic resolution, column life time and sample preparation. Due to the simplified analytical workflow of the analytical procedure the application's integration into the collection of methods of interested laboratories should be facilitated.
Full text links
Trending Papers
Acute and non-acute decompensation of liver cirrhosis (47/130).Liver International : Official Journal of the International Association for the Study of the Liver 2024 March 2
Guide to Utilization of the Microbiology Laboratory for Diagnosis of Infectious Diseases: 2024 Update by the Infectious Diseases Society of America (IDSA) and the American Society for Microbiology (ASM).Clinical Infectious Diseases 2024 March 6
Ten Influential Point-of-Care Ultrasound Papers: 2023 in Review.Journal of Intensive Care Medicine 2024 Februrary 20
Administration of methylene blue in septic shock: pros and cons.Critical Care : the Official Journal of the Critical Care Forum 2024 Februrary 17
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app