Impact of the calibration conditions of XR-RV3 films on peak skin dose measurements in interventional radiology.

In this study, the impact of radiochromic films' (XR-RV3) calibration on PSD measurements was investigated under various peak kilovoltage (kVp) and additional filtration conditions. Films were calibrated free-in-air for six beam qualities with Allura Xper FD20 system (Philips). Six calibration curves (CCs) were constructed. Each beam quality was characterized in terms of mean energy (ME) in the air, with table, with table and water phantom using Monte Carlo simulations. A cohort of 155 patient films from cardiology (37) and vascular (118) procedures were read with each CC. Routine calibration beam quality was taken as reference (DoseNorm). Overall, it was observed that for a wider ME difference between the exposed film and the CC used, a larger deviation (from -28% to +41%) was observed. The choice of beam quality for the calibration is a key point when additional filtration and kVp are automatically controlled in clinical conditions.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app