OBJECTIVE: We aimed to elucidate the effects of the micro-structure of the pyrolytic carbon for artificial heart valves on its hydrodynamic performance.

METHODS: Bileaflet mechanical valves of GKS 23 and 29 A were randomly selected. According to ISO5840, mean transvalvular pressure (MPG), regurgitation fraction (RF), and effective orifice area (EOA) of valve were assessed. Then, parallel-groove pattern was constructed by laser etching on leaflet surface, and the valves were subjected again to the same test.

RESULTS: Compared with before patterning at 2, 3.5, 5, and 7 L/min, the MPG of the valves in two specifications were higher, the EOA was larger in 23 A, but smaller in 29 A, and the RF was contrary to EOA. At 5 L/min, the RF in both specifications was lower after etching at 45 bpm. At 70 bpm however, the RF in 23 A decreased, in 29 A increased.

CONCLUSION: The parallel-groove pattern on leaflet surface affected the hemodynamic performance of the valve prostheses.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app