Single-use versus reusable metallic laryngoscopes for non-emergent intubation: A retrospective review of 72,672 intubations.

STUDY OBJECTIVE: Increased regulatory requirements for sterilization in recent years have prompted a widespread transition from reusable to single-use laryngoscopes. The purpose of this study was to determine if the transition from metallic reusable to metallic single-use laryngoscopes impacted the performance of direct laryngoscopy at an academic medical center.

DESIGN: Single-site retrospective cohort study.

SETTING: General anesthetic cases requiring tracheal intubation.

PATIENTS: Adult patients undergoing non-emergent procedures.

INTERVENTIONS: Data were collected two years before and two years after a transition from metallic reusable to metallic single-use laryngoscopes.

MEASUREMENTS: The primary outcome was need for intubation rescue with an alternate device. Secondary outcomes were difficult laryngeal view (modified Cormack-Lehane grade ≥ 2b) and hypoxemia (SpO2  < 90% for >30 s) during direct laryngoscopy intubations. Subgroup analyses for rapid sequence induction, Macintosh blades, Miller blades, and patients with difficult airway risk factors (Obstructive Sleep Apnea, Mallampati ≥3, Body Mass Index >30 kg/m2 ) were performed.

MAIN RESULTS: In total, 72,672 patients were included: 35,549 (48.9%) in the reusable laryngoscope cohort and 37,123 (51.1%) in the single-use laryngoscope cohort. Compared with reusable laryngoscopes, single-use laryngoscopes were associated with fewer rescue intubations with an alternate device (covariates-adjusted odds ratio [OR] 0.81 95% CI 0.66-0.99). Single-use laryngoscopes were also associated with lower odds of difficult laryngeal view (OR 0.86; 95% CI 0.80-0.93). Single use laryngoscopes were not associated with hypoxemia during the intubation attempt (OR 1.03; 95% CI 0.88-1.20). Similar results were observed for subgroup analyses including rapid sequence induction, Macintosh blades, Miller blades, and patients with difficult airway risk factors.

CONCLUSIONS: Metallic single-use laryngoscopes were associated with less need for rescue intubation with alternate devices and lower incidence of poor laryngeal view compared to reusable metallic laryngoscopes.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app