How and when to measure pH in IVF culture media: validation of a portable blood gas analyzer in two IVF culture dishes for time lapse and conventional incubators.
PURPOSE: Maintaining a stable pH at optimal level in human embryo culture media is crucial for embryo development but poses a challenge for all IVF laboratories. We validate analytically reliable conditions for pH measurement that are as close as possible to the embryo microenvironment during IVF.
METHODS: This was a multicentric study. A Siemens EPOC portable blood gas analyzer was used. The analytical validation was carried out under the culture medium (Global Total HSA®) conditions of use (microdroplets, under oil overlay, in a IVF incubator with (EmbryoScope®) or without a time lapse system (K system G210+®) and using IVF dishes. The validation included repeatability ("within-run" precision), total precision (between-day precision), trueness based on inter-laboratory comparison, inaccuracy based on external quality assessment and comparison to the reference technique. We also assessed the pre-analytical medium incubation time required to obtain a target value.
RESULTS: A measurement after an incubation period of 24 to 48 h is more representative of the pH to which the embryo will be exposed throughout the culture. The "within-run" and "between-day" precision show very low coefficients of variation (CV%): 0.17 to 0.22% and 0.13 to 0.34%, respectively, with IVF culture media. Trueness (% bias) range from - 0.07 to - 0.03%. We demonstrate good correlation between EPOC and reference pH electrode with an overestimation of 0.03 pH units of EPOC.
CONCLUSION: Our method demonstrates good analytical performance for IVF laboratories wishing to implement a robust quality assurance system to monitor pH in embryo culture media. Compliance with stringent pre-analytical and analytical conditions is essential.
METHODS: This was a multicentric study. A Siemens EPOC portable blood gas analyzer was used. The analytical validation was carried out under the culture medium (Global Total HSA®) conditions of use (microdroplets, under oil overlay, in a IVF incubator with (EmbryoScope®) or without a time lapse system (K system G210+®) and using IVF dishes. The validation included repeatability ("within-run" precision), total precision (between-day precision), trueness based on inter-laboratory comparison, inaccuracy based on external quality assessment and comparison to the reference technique. We also assessed the pre-analytical medium incubation time required to obtain a target value.
RESULTS: A measurement after an incubation period of 24 to 48 h is more representative of the pH to which the embryo will be exposed throughout the culture. The "within-run" and "between-day" precision show very low coefficients of variation (CV%): 0.17 to 0.22% and 0.13 to 0.34%, respectively, with IVF culture media. Trueness (% bias) range from - 0.07 to - 0.03%. We demonstrate good correlation between EPOC and reference pH electrode with an overestimation of 0.03 pH units of EPOC.
CONCLUSION: Our method demonstrates good analytical performance for IVF laboratories wishing to implement a robust quality assurance system to monitor pH in embryo culture media. Compliance with stringent pre-analytical and analytical conditions is essential.
Full text links
Related Resources
Trending Papers
Consensus Statement on Vitamin D Status Assessment and Supplementation: Whys, Whens, and Hows.Endocrine Reviews 2024 April 28
The Tricuspid Valve: A Review of Pathology, Imaging, and Current Treatment Options: A Scientific Statement From the American Heart Association.Circulation 2024 April 26
British Society of Gastroenterology guidelines for the management of hepatocellular carcinoma in adults.Gut 2024 April 17
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Ventilator Waveforms May Give Clues to Expiratory Muscle Activity.American Journal of Respiratory and Critical Care Medicine 2024 April 25
Systemic lupus erythematosus.Lancet 2024 April 18
Acute Kidney Injury and Electrolyte Imbalances Caused by Dapagliflozin Short-Term Use.Pharmaceuticals 2024 March 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app