Accuracy and Reliability of Whole Blood Bilirubin Measurements Using a Roche Blood Gas Analyzer for Neonatal Hyperbilirubinemia Screening and Risk Stratification.
Background: Accurate bilirubin measurements are essential for appropriate management of neonatal hyperbilirubinemia. This study aimed to evaluate the accuracy and reliability of whole blood bilirubin measurements obtained using a Roche blood gas analyzer (Roche TBiL), with total serum bilirubin (TSB) measurements determined by the Ortho VITROS 4600 chemistry system (Ortho TSB) serving as a reference.
Materials and Methods: Medical records of hospitalized neonates that underwent simultaneous Roche TBiL and Ortho TSB measurements were reviewed for eligibility selection and data collection. The correlations and differences between two sets of results were determined using Passing-Bablok regression analysis and a Bland-Altman plot, respectively. For eligible newborns, the risk of developing severe hyperbilirubinemia was assessed using the Bhutani nomogram. Weighted kappa analysis was used to evaluate the agreement between risk prediction by the two methods.
Results: We obtained 618 paired Roche TBiL and Ortho TSB results from 309 neonates. Roche TBiL and Ortho TSB measurements showed a good correlation ( r = 0.923; 95% CI: 0.905-0.938). Passing-Bablok regression analysis yielded the following equation: Roche TBiL = 0.794 × Ortho TSB + 1.255 mg/dL, with a slope of 0.794 (95% CI: 0.763-0.825) and intercept of 1.255 (95% CI: 1.042-1.417). The average difference between the two methods was 0.1 ± 1.448 mg/dL. A total of 207 neonates were eligible for evaluation of the agreement between the risk-grading methods. Although kappa analysis showed good agreement between the methods, with a weighted kappa of 0.681 (95% CI: 0.610-0.751) across all populations, the values for approximately half of the neonates at intermediate and high risk of hyperbilirubinemia (33/72) were underestimated by Roche TBiL.
Conclusion: Our results indicate that Roche TBiL and Ortho TSB measurements in the neonatal population are not consistent. As a point-of-care and trace blood assay, Roche blood gas bilirubin measurements can facilitate primary screening of neonatal hyperbilirubinemia, but it seems to lack accuracy regarding risk stratification, particularly for high-risk newborn individuals.
Materials and Methods: Medical records of hospitalized neonates that underwent simultaneous Roche TBiL and Ortho TSB measurements were reviewed for eligibility selection and data collection. The correlations and differences between two sets of results were determined using Passing-Bablok regression analysis and a Bland-Altman plot, respectively. For eligible newborns, the risk of developing severe hyperbilirubinemia was assessed using the Bhutani nomogram. Weighted kappa analysis was used to evaluate the agreement between risk prediction by the two methods.
Results: We obtained 618 paired Roche TBiL and Ortho TSB results from 309 neonates. Roche TBiL and Ortho TSB measurements showed a good correlation ( r = 0.923; 95% CI: 0.905-0.938). Passing-Bablok regression analysis yielded the following equation: Roche TBiL = 0.794 × Ortho TSB + 1.255 mg/dL, with a slope of 0.794 (95% CI: 0.763-0.825) and intercept of 1.255 (95% CI: 1.042-1.417). The average difference between the two methods was 0.1 ± 1.448 mg/dL. A total of 207 neonates were eligible for evaluation of the agreement between the risk-grading methods. Although kappa analysis showed good agreement between the methods, with a weighted kappa of 0.681 (95% CI: 0.610-0.751) across all populations, the values for approximately half of the neonates at intermediate and high risk of hyperbilirubinemia (33/72) were underestimated by Roche TBiL.
Conclusion: Our results indicate that Roche TBiL and Ortho TSB measurements in the neonatal population are not consistent. As a point-of-care and trace blood assay, Roche blood gas bilirubin measurements can facilitate primary screening of neonatal hyperbilirubinemia, but it seems to lack accuracy regarding risk stratification, particularly for high-risk newborn individuals.
Full text links
Related Resources
Trending Papers
Consensus Statement on Vitamin D Status Assessment and Supplementation: Whys, Whens, and Hows.Endocrine Reviews 2024 April 28
The Tricuspid Valve: A Review of Pathology, Imaging, and Current Treatment Options: A Scientific Statement From the American Heart Association.Circulation 2024 April 26
British Society of Gastroenterology guidelines for the management of hepatocellular carcinoma in adults.Gut 2024 April 17
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Ventilator Waveforms May Give Clues to Expiratory Muscle Activity.American Journal of Respiratory and Critical Care Medicine 2024 April 25
Systemic lupus erythematosus.Lancet 2024 April 18
Acute Kidney Injury and Electrolyte Imbalances Caused by Dapagliflozin Short-Term Use.Pharmaceuticals 2024 March 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app