Pleural fluid glucose testing using a finger stick glucometer: a novel bedside test.
Journal of Thoracic Disease 2019 November
BACKGROUND: Pleural fluid glucose (PFG) has diagnostic and therapeutic implications for the management of pleural effusion. The literature examining point-of-care testing of PFG is limited, and no studies exist for the bedside measurement of PFG using a glucometer (B-PFG). In this study, we compared the accuracy of B-PFG measurement to standard in-lab measurement (Lab-PFG).
METHODS: Patients undergoing thoracentesis or thoracostomy were enrolled. PFG was measured at the bedside with a finger stick blood glucometer (ACCU-CHEK® Inform II, Roche) and in the laboratory. Two consecutive measurements were taken using the glucometer, and their mean was compared to the glucose concentration measured in the laboratory. Pearson correlation coefficient and Bland-Altman Plot analysis were used to compare the two measurements.
RESULTS: Sixty patients were included. Mean age was 64.1 years. Forty-nine patients had exudative effusions (41% malignant, 26% parapneumonic, and 33% others). There was a significant correlation between the B-PFG and the Lab-PFG (r=0.98, 95% CI of 0.97 to 0.99; P<0.0001). There was good agreement between the B-PFG and the Lab-PFG with a mean difference of 14.8 mg/dL [95% limit of agreement (LOA) of -2.2 to 31.8 mg/dL]. This agreement was even better at glucose values less than 80 mg/dL.
CONCLUSIONS: PFG measured at the bedside with a glucometer closely correlates with the laboratory measurement. Further studies are needed prior to incorporating this test in clinical practice.
METHODS: Patients undergoing thoracentesis or thoracostomy were enrolled. PFG was measured at the bedside with a finger stick blood glucometer (ACCU-CHEK® Inform II, Roche) and in the laboratory. Two consecutive measurements were taken using the glucometer, and their mean was compared to the glucose concentration measured in the laboratory. Pearson correlation coefficient and Bland-Altman Plot analysis were used to compare the two measurements.
RESULTS: Sixty patients were included. Mean age was 64.1 years. Forty-nine patients had exudative effusions (41% malignant, 26% parapneumonic, and 33% others). There was a significant correlation between the B-PFG and the Lab-PFG (r=0.98, 95% CI of 0.97 to 0.99; P<0.0001). There was good agreement between the B-PFG and the Lab-PFG with a mean difference of 14.8 mg/dL [95% limit of agreement (LOA) of -2.2 to 31.8 mg/dL]. This agreement was even better at glucose values less than 80 mg/dL.
CONCLUSIONS: PFG measured at the bedside with a glucometer closely correlates with the laboratory measurement. Further studies are needed prior to incorporating this test in clinical practice.
Full text links
Related Resources
Trending Papers
Guillain-Barré syndrome: History, pathogenesis, treatment, and future directions.European Journal of Neurology 2024 May 17
Contrast-induced acute kidney injury: a review of definition, pathogenesis, risk factors, prevention and treatment.BMC Nephrology 2024 April 23
Angiotensin Receptor Blocker-Neprilysin Inhibitor for Heart Failure with Reduced Ejection Fraction.Pharmacological Research : the Official Journal of the Italian Pharmacological Society 2024 May 12
The Therapy and Management of Heart Failure with Preserved Ejection Fraction: New Insights on Treatment.Cardiac Failure Review 2024
European Respiratory Society Clinical Practice Guideline on symptom management for adults with serious respiratory illness.European Respiratory Journal 2024 May 9
Axillary Surgery for Breast Cancer in 2024.Cancers 2024 April 24
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app