We have located links that may give you full text access.
On the accuracy of cochlear duct length measurement in computed tomographic images.
PURPOSE: Patient specific selection of cochlear implants would benefit from pre-operative knowledge of cochlear length. Several methods for its measurement or estimation have been described in literature. This study focused on the achievable accuracy in clinically available imaging.
METHODS: Five simplified cochlea models milled into porcine bone were scanned in water using clinical cone beam computed tomography. Due to their well-known dimensions these phantoms served as gold standard for the length measurements. Each phantom was measured ten times using the custom software Comet. In addition, cochleae in ten image datasets taken indiscriminately from clinical routine were measured ten times each to test the precision under realistic conditions. The results were also compared to estimations based on the diameter of the basal turn (A value) as described in literature.
RESULTS: Measurement accuracy of the phantoms' lengths was high (average error: - 0.2 mm; standard deviation: 0.3 mm). The pooled standard deviation for the measurements in clinical datasets was 0.6 mm. Errors resulted mainly from problems locating the helicotrema. The estimations differed on average - 1.7 to + 0.4 mm from the manual measurements and had standard deviations between 0.5 and 0.6 mm depending on the algorithm.
CONCLUSIONS: The program Comet was successfully used to accurately measure the length of the cochlea models in clinically available imaging. The lower image quality of patient scans reduced the precision of the measurement. Estimations using the A value are a quicker alternative for averagely sized cochleae in cases where the lack of accuracy is tolerable.
METHODS: Five simplified cochlea models milled into porcine bone were scanned in water using clinical cone beam computed tomography. Due to their well-known dimensions these phantoms served as gold standard for the length measurements. Each phantom was measured ten times using the custom software Comet. In addition, cochleae in ten image datasets taken indiscriminately from clinical routine were measured ten times each to test the precision under realistic conditions. The results were also compared to estimations based on the diameter of the basal turn (A value) as described in literature.
RESULTS: Measurement accuracy of the phantoms' lengths was high (average error: - 0.2 mm; standard deviation: 0.3 mm). The pooled standard deviation for the measurements in clinical datasets was 0.6 mm. Errors resulted mainly from problems locating the helicotrema. The estimations differed on average - 1.7 to + 0.4 mm from the manual measurements and had standard deviations between 0.5 and 0.6 mm depending on the algorithm.
CONCLUSIONS: The program Comet was successfully used to accurately measure the length of the cochlea models in clinically available imaging. The lower image quality of patient scans reduced the precision of the measurement. Estimations using the A value are a quicker alternative for averagely sized cochleae in cases where the lack of accuracy is tolerable.
Full text links
Related Resources
Trending Papers
Consensus Statement on Vitamin D Status Assessment and Supplementation: Whys, Whens, and Hows.Endocrine Reviews 2024 April 28
The Tricuspid Valve: A Review of Pathology, Imaging, and Current Treatment Options: A Scientific Statement From the American Heart Association.Circulation 2024 April 26
British Society of Gastroenterology guidelines for the management of hepatocellular carcinoma in adults.Gut 2024 April 17
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Ventilator Waveforms May Give Clues to Expiratory Muscle Activity.American Journal of Respiratory and Critical Care Medicine 2024 April 25
Systemic lupus erythematosus.Lancet 2024 April 18
Acute Kidney Injury and Electrolyte Imbalances Caused by Dapagliflozin Short-Term Use.Pharmaceuticals 2024 March 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app