Foreign bodies in small arteries after use of an infusion microcatheter.

Over a 31-month period, we performed four neurointerventional procedures after which unexpected foreign bodies were noted in multiple arteries. All four procedures had in common the use of Fastracker-18 infusion microcatheters. Histologically, the intravascular debris looked strikingly similar to the hydrophilic coating on the catheter. An in vitro test mimicking clinical use of the microcatheter revealed that the hydrophilic coating can separate from the catheter. Until the coating is refined to make it more resistant to stripping, it may be advisable to reduce the amount of back-and-forth movement of these microcatheters if they have been positioned through guide catheters with small inner diameters and angled tips.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app