Adaptive higher-order singular value decomposition clutter filter for ultrafast Doppler imaging of coronary flow under non-negligible tissue motion.

Ultrasonics 2024 March 27
BACKGROUND AND OBJECTIVE: With the development of advanced clutter-filtering techniques by singular value decomposition (SVD) and leveraging favorable acquisition settings such as open-chest imaging by a linear high-frequency probe and plane waves, several studies have shown the feasibility of cardiac flow measurements during the entire cardiac cycle, ranging from coronary flow to myocardial perfusion. When applying these techniques in a routine clinical setting, using transthoracic ultrasound imaging, new challenges emerge. Firstly, a smaller aperture is needed that can fit between ribs. Consequently, diverging waves are employed instead of plane waves to achieve an adequate field of view. Secondly, to ensure imaging at a larger depth, the maximum pulse repetition frequency has to be reduced. Lastly, in comparison to the open-chest scenario, tissue motion induced by the heartbeat is significantly stronger. The latter complicates substantially the distinction between clutter and blood signals.

METHODS: This study investigates a strategy to overcome these challenges by diverging wave imaging with an optimal number of tilt angles, in combination with dedicated clutter-filtering techniques. In particular, a novel, adaptive, higher-order SVD (HOSVD) clutter filter, which utilizes spatial, temporal, and angular information of the received ultrasound signals, is proposed to enhance clutter and blood separation.

RESULTS: When non-negligible tissue motion is present, using fewer tilt angles not only reduces the decorrelation between the received waveforms but also allows for collecting more temporal samples at a given ensemble duration, contributing to improved Doppler performance. The addition of a third angular dimension enables the application of HOSVD, providing greater flexibility in selecting blood separation thresholds from a 3-D tensor. This differs from the conventional threshold selection method in a 2-D spatiotemporal space using SVD. Exhaustive threshold search has shown a significant improvement in Contrast and Contrast-to-Noise ratio for Power Doppler images filtered with HOSVD compared to the SVD-based clutter filter.

CONCLUSION: With the improved settings, the obtained Power Doppler images show the feasibility of measuring coronary flow under the influence of non-negligible tissue motion in both in vitro and ex vivo.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app