Characterization of Corneal Biomechanics Using CORVIS ST Device in Different Grades of Myopia in a Sample of Middle Eastern Ethnicity.
PURPOSE: To characterize corneal biomechanical properties using the CORVIS-ST device in myopic individuals.
METHODS: This prospective cross-sectional study included patients with myopia. Our study included 154 eyes of 154 myopic patients aged between 18 and 40 years, with stable refraction for at least 2 years. A full ophthalmological examination and corneal tomography were performed using a Pentacam HR device. Corneal biomechanical parameters were assessed using the CORVIS-ST device in mild, moderate, severe, and extreme myopia groups.
RESULTS: Statistically significant differences were observed in the DA ratio (p = 0.033), SP-A (p=0.009), CBI (p=0.041), SSI (p=0.000), and Peak distance (p = 0.032). In correlation with different Corvis ST biomechanical variables, SE was found to be correlated with DA ratio(r=-0.191, p=0.018), SP-A(r=0.199, p=0.013) and SSI(r=-0.336, p=0.000), while in multiple regression analysis, SE was found to be independently correlated with SSI and peak distance(p=0.036,0.038 respectively) while the grade of myopia was found to be independently correlated with SP-A(p=0.034).
CONCLUSION: SSI, Peak distance, and SP-A were independently related to SE and myopia grade, confirming the hypothesis that eyes with higher myopia are more deformable and less stress resistant.
METHODS: This prospective cross-sectional study included patients with myopia. Our study included 154 eyes of 154 myopic patients aged between 18 and 40 years, with stable refraction for at least 2 years. A full ophthalmological examination and corneal tomography were performed using a Pentacam HR device. Corneal biomechanical parameters were assessed using the CORVIS-ST device in mild, moderate, severe, and extreme myopia groups.
RESULTS: Statistically significant differences were observed in the DA ratio (p = 0.033), SP-A (p=0.009), CBI (p=0.041), SSI (p=0.000), and Peak distance (p = 0.032). In correlation with different Corvis ST biomechanical variables, SE was found to be correlated with DA ratio(r=-0.191, p=0.018), SP-A(r=0.199, p=0.013) and SSI(r=-0.336, p=0.000), while in multiple regression analysis, SE was found to be independently correlated with SSI and peak distance(p=0.036,0.038 respectively) while the grade of myopia was found to be independently correlated with SP-A(p=0.034).
CONCLUSION: SSI, Peak distance, and SP-A were independently related to SE and myopia grade, confirming the hypothesis that eyes with higher myopia are more deformable and less stress resistant.
Full text links
Related Resources
Trending Papers
Renin-Angiotensin-Aldosterone System: From History to Practice of a Secular Topic.International Journal of Molecular Sciences 2024 April 5
Prevention and treatment of ischaemic and haemorrhagic stroke in people with diabetes mellitus: a focus on glucose control and comorbidities.Diabetologia 2024 April 17
British Society for Rheumatology guideline on management of adult and juvenile onset Sjögren disease.Rheumatology 2024 April 17
Albumin: a comprehensive review and practical guideline for clinical use.European Journal of Clinical Pharmacology 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app