Choosing Optimal Cutoff Frequencies for Filtering Linear Acceleration and Angular Velocity Signals Associated with Head Impacts Measured by Instrumented Mouthguards.
Annals of Biomedical Engineering 2024 Februrary 26
Head impact sensors worn in the mouth are popular because they couple directly to the teeth and provide six-degree-of-freedom head measurements. Mouthpiece signal filters have conventionally used cutoff frequencies lower than recommended practices (Society of Automotive Engineers, SAE J211-1) to eliminate extraneous noise when measuring with live subjects. However, there is little information about the effects of filter choice on the accuracy of signals measured by instrumented mouthpieces. Lack of standardization in head impact measurement device post-processing techniques can result in data that are not comparable across studies or device brands. This study sought optimal filter cutoff frequencies for six-degree-of-freedom measurements made at the teeth using instrumented mouthguards. We collected linear acceleration and angular velocity signals at the head center of gravity (CG) using laboratory-grade instrumentation. We also collected and filtered similar six-degree-of-freedom measurements from an instrumented mouthguard using 24 cutoff frequencies, from 25 to 600 Hz. We transformed the measurements to linear acceleration at the center of gravity of the head (CG) using all kinematic variables at the teeth, optimizing linear and angular mouthguard cutoff frequencies with one equation. We calculated the percent error in transformed peak resultant linear acceleration and minimized the mean and standard deviation in error. The optimal cutoff frequencies were 175 Hz for linear acceleration and 250 Hz for angular velocity. Rigid impacts (3-5 ms duration) had higher optimal cutoff frequencies (175 Hz linear acceleration, 275 Hz angular velocity) than padded impacts (10-12 ms duration; 100 Hz linear acceleration, 175 Hz angular velocity), and all impacts together (3-12 ms duration; 175 Hz linear acceleration, 250 Hz angular velocity). Instrumented mouthpiece manufacturers and researchers using these devices should consider these optimal filter cutoff frequencies to minimize measurement error. Sport-specific filter criteria for teeth-based sensors may be warranted to account for the difference in optimal cutoff frequency combination by impact duration.
Full text links
Related Resources
Trending Papers
Renin-Angiotensin-Aldosterone System: From History to Practice of a Secular Topic.International Journal of Molecular Sciences 2024 April 5
Prevention and treatment of ischaemic and haemorrhagic stroke in people with diabetes mellitus: a focus on glucose control and comorbidities.Diabetologia 2024 April 17
British Society for Rheumatology guideline on management of adult and juvenile onset Sjögren disease.Rheumatology 2024 April 17
Albumin: a comprehensive review and practical guideline for clinical use.European Journal of Clinical Pharmacology 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app