Correlation between the electrocardiogram amplitude detected by an implantable cardiac monitor and the implantation depth.
Annals of Noninvasive Electrocardiology 2023 December 14
INTRODUCTION: Implantable cardiac monitors (ICMs) primarily use R-R intervals in subcutaneous electrocardiograms (ECGs) to detect arrhythmias. Therefore, reliable detection of R-wave amplitude by an ICM is vital. Since ICMs detect subcutaneous ECGs, the impact of the implantation depth should be assessed.
METHODS AND RESULTS: This study investigated the influence of ICM depth on R-wave (ICM-R) amplitude on an ECG generated by an ICM (JOT Dx; Abbott). Overall, 53 patients who underwent ICM implantation at Kamagaya General Hospital from May 2022 to April 2023 were retrospectively reviewed. The depth-position was measured using ultrasound imaging after implantation. The depth of the ICM did not show any correlation with ICM-R amplitude (r = -.0141, p = .294). However, the distance between the ICM and the heart surface showed a significant correlation with ICM-R amplitude (r = -.581, p < .001). Body weight (r = -.0283, p = .033) and body mass index (r = -.0342, p = .009) were associated with ICM-R amplitude. S wave in the V1 -lead was also associated with ICM-R amplitude (r = .481, p < .001). After multivariate analysis, the distance between the ICM and heart surface and the S wave in V1 were independent determinants for the ICM-R amplitude.
CONCLUSION: The ICM-R amplitude may be higher with the ICM implanted deeper.
METHODS AND RESULTS: This study investigated the influence of ICM depth on R-wave (ICM-R) amplitude on an ECG generated by an ICM (JOT Dx; Abbott). Overall, 53 patients who underwent ICM implantation at Kamagaya General Hospital from May 2022 to April 2023 were retrospectively reviewed. The depth-position was measured using ultrasound imaging after implantation. The depth of the ICM did not show any correlation with ICM-R amplitude (r = -.0141, p = .294). However, the distance between the ICM and the heart surface showed a significant correlation with ICM-R amplitude (r = -.581, p < .001). Body weight (r = -.0283, p = .033) and body mass index (r = -.0342, p = .009) were associated with ICM-R amplitude. S wave in the V1 -lead was also associated with ICM-R amplitude (r = .481, p < .001). After multivariate analysis, the distance between the ICM and heart surface and the S wave in V1 were independent determinants for the ICM-R amplitude.
CONCLUSION: The ICM-R amplitude may be higher with the ICM implanted deeper.
Full text links
Related Resources
Trending Papers
Revascularization Strategy in Myocardial Infarction with Multivessel Disease.Journal of Clinical Medicine 2024 March 27
Intravenous infusion of dexmedetomidine during the surgery to prevent postoperative delirium and postoperative cognitive dysfunction undergoing non-cardiac surgery: a meta-analysis of randomized controlled trials.European Journal of Medical Research 2024 April 19
The Tricuspid Valve: A Review of Pathology, Imaging, and Current Treatment Options: A Scientific Statement From the American Heart Association.Circulation 2024 April 26
Consensus Statement on Vitamin D Status Assessment and Supplementation: Whys, Whens, and Hows.Endocrine Reviews 2024 April 28
Management of Diverticulitis: A Review.JAMA Surgery 2024 April 18
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app