Total Reference Air Kerma (TRAK) is Associated with Dosimetric Parameters in Template-Based High Dose-Rate (HDR) Interstitial Brachytherapy in Advanced Gynecologic Cancers.

PURPOSE/OBJECTIVE(S): To study TRAK and its association with dosimetric parameters in template-based high dose-rate interstitial brachytherapy in advanced gynecologic cancers.

MATERIALS/METHODS: Brachytherapy treatment plans of 53 patients treated between 2012 and 2022 at our institution with template-based Iridium-192 HDR brachytherapy, post-external beam RT, for locally advanced cancers of the cervix and vagina were retrospectively reviewed. Brachytherapy dose ranged from 25 to 30-Gy delivered in 4 to 6 fractions. The median number of flexi-guide catheters implanted was 18 (range 10-30). Clinical Target Volume (CTV) values were mean (±SD): 72.2 (±40.4) cm3 (high-risk, HR) and 182.2 (±73.7) cm3 (intermediate-risk, IR) respectively. TRAK per fraction (cGy at 1m), dose-volume information for the implant, target, and organ-at-risk (OAR) were recorded. Indices for dose coverage (CI), homogeneity (DHI), non-uniformity (DNR), overdose volume (ODI) were computed. Regression and correlation tests were used to study the TRAK relationship with various dosimetric parameters. The false discovery rate at a 5% level was corrected using the Benjamini-Hochberg procedure.

RESULTS: The average TRAK per fraction was 0.365 (±0.12) cGy. Mean and range values of plan quality indices were - CI 0.92 (0.7- 1.0), DHI 0.57 (0.41 - 0.77), DNR 0.43 (0.23 - 0.59) and ODI 0.22 (0.11 - 0.38), respectively Correlation results for TRAK with various dosimetric indices are presented in Table 1. TRAK showed a weak correlation with the number of flexi-guide catheters implanted (r = 0.35, p = 0.013). TRAK correlated strongly with target volumes (CTV_HR and CTV_IR and CTV_HR V100% ) and with isodose volumes at both high (V300 , V200 , V150 ), and low dose levels (V90 , V85 , V50 ) (p<0.00001). TRAK correlated moderately with OAR 2-cm3 doses (p<0.00001). A weak correlation was observed between TRAK and plan quality indices.

CONCLUSION: TRAK correlates positively with target volume and volumes enclosed by isodoses at various dose levels in interstitial HDR brachytherapy of advanced Gynecologic Cancers. Interestingly, our study observed a comparatively stronger positive correlation between TRAK and Sigmoid 2-cm3 dose, equated to TRAK correlation with bladder, rectum, and small bowel 2-cm3 doses. This finding could interest future studies utilizing TRAK as a surrogate for treatment outcome and toxicity.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app