Real-World Utilization of an Intrauterine, Vacuum-Induced, Hemorrhage-Control Device.

Obstetrics and Gynecology 2023 September 15
OBJECTIVE: To assess the real-world effectiveness and safety of a U.S. Food and Drug Administration (FDA)-cleared intrauterine vacuum-induced-hemorrhage control device for postpartum hemorrhage (PPH) management.

METHODS: Sixteen centers in the United States participated in this observational, postmarket registry medical record review (October 2020 through March 2022). The primary effectiveness outcome was treatment success, defined as bleeding control after insertion with no treatment escalation or bleeding recurrence. Additional outcomes included blood loss, time to device insertion, indwelling time, bleeding recurrence, and time to bleeding control. Treatment success and severe maternal morbidity measures (transfusion of 4 or more units of red blood cell, intensive care unit admission, and hysterectomy) were evaluated by blood loss before insertion. To assess safety, serious adverse events (SAEs) and adverse device effects were collected. All outcomes were summarized by mode of delivery; treatment success was summarized by bleeding cause (all causes, any atony, isolated atony, nonatony).

RESULTS: In total, 800 individuals (530 vaginal births, 270 cesarean births) were treated with the device; 94.3% had uterine atony (alone or in combination with other causes). Median total blood loss at device insertion was 1,050 mL in vaginal births and 1,600 mL in cesarean births. Across all bleeding causes, the treatment success rate was 92.5% for vaginal births and was 83.7% for cesarean births (95.8% [n=307] and 88.2% [n=220], respectively, in isolated atony). Median indwelling time was 3.1 hours and 4.6 hours, respectively. In vaginal births, 14 SAEs were reported among 13 individuals (2.5%). In cesarean births, 22 SAEs were reported among 21 individuals (7.8%). Three (0.4%) SAEs were deemed possibly related to the device or procedure. No uterine perforations or deaths were reported.

CONCLUSION: For both vaginal and cesarean births in real-world settings, rapid and effective bleeding control was achieved with an FDA-cleared intrauterine vacuum-induced hemorrhage-control device. The safety profile was consistent with that observed in the registrational trial (NCT02883673), and SAEs or adverse device effects were of the nature and severity expected in the setting of PPH. This device is an important new tool for managing a life-threatening condition, and timely utilization may help to improve obstetric hemorrhage outcomes.

CLINICAL TRIAL REGISTRATION: ClinicalTrials.gov, NCT04995887.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app