A Tool to Integrate Electrophysiological Mapping for Cardiac Radioablation of Ventricular Tachycardia.

PURPOSE: Cardiac radioablation is an emerging therapy for recurrent ventricular tachycardia. Electrophysiology (EP) data, including electroanatomic maps (EAM) and electrocardiographic imaging (ECGI), provide crucial information for defining the arrhythmogenic target volume. The absence of standardized workflows and software tools to integrate the EP maps into a radiation planning system limits their use. This study developed a comprehensive software tool to enable efficient utilization of the mapping for cardiac radioablation treatment planning.

METHODS AND MATERIALS: The tool, HeaRTmap, is a Python-scripted plug-in module on the open-source 3D Slicer software platform. HeaRTmap is able to import EAM and ECGI data and visualize the maps in 3D Slicer. The EAM is translated into a 3D space by registration with cardiac magnetic resonance images (MRI) or computed tomography (CT) . After the scar area is outlined on the mapping surface, the tool extracts and extends the annotated patch into a closed surface and converts it into a structure set associated with the anatomic images. The tool then exports the structure set and the images as The Digital Imaging and Communications in Medicine Standard in Radiotherapy for a radiation treatment planning system to import. Overlapping the scar structure on simulation CT, a transmural target volume is delineated for treatment planning.

RESULTS: The tool has been used to transfer Ensite NavX EAM data into the Varian Eclipse treatment planning system in radioablation on 2 patients with ventricular tachycardia. The ECGI data from CardioInsight was retrospectively evaluated using the tool to derive the target volume for a patient with left ventricular assist device, showing volumetric matching with the clinically used target with a Dice coefficient of 0.71.

CONCLUSIONS: HeaRTmap smoothly fuses EP information from different mapping systems with simulation CT for accurate definition of radiation target volume. The efficient integration of EP data into treatment planning potentially facilitates the study and adoption of the technique.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app