Subcutaneous implantable cardioverter-defibrillator noise following left ventricular assist device implantation.
Journal of Arrhythmia 2023 April
BACKGROUND: The incidence and impact of noise in a subcutaneous implantable cardioverter defibrillator (S-ICD) after left ventricular assist device (LVAD) implantation is not well established.
METHODS: We performed a retrospective study of patients implanted with LVAD and with a pre-existing S-ICD between January 2005 and December 2020 at the three Mayo Clinic centers (Minnesota, Arizona, and Florida).
RESULTS: Of the 908 LVAD patients, a pre-existing S-ICD was present in 9 patients (mean age 49.1 ± 13.7 years, 66.7% males), 100% with Boston Scientific third-generation EMBLEM MRI S-ICD, 11% with HeartMate II (HM II), 44% with HeartMate 3 (HM 3), and 44% with HeartWare (HW) LVAD. The incidence of noise from LVAD-related electromagnetic interference (EMI) was 33% and was only seen with HM 3 LVAD. Multiple measures attempted to resolve noise, including using alternative S-ICD sensing vector, adjusting S-ICD time zone, and increasing LVAD pump speed, were unsuccessful, necessitating S-ICD device therapies to be turned off permanently.
CONCLUSIONS: The incidence of LVAD-related S-ICD noise is high in patients with concomitant LVAD and S-ICD with significant impact on device function. As conservative management failed to resolve the EMI, the S-ICDs had to be programmed off to avoid inappropriate shocks. This study highlights the importance of awareness of LVAD-SICD device interference and the need to improve S-ICD detection algorithms to eliminate noise.
METHODS: We performed a retrospective study of patients implanted with LVAD and with a pre-existing S-ICD between January 2005 and December 2020 at the three Mayo Clinic centers (Minnesota, Arizona, and Florida).
RESULTS: Of the 908 LVAD patients, a pre-existing S-ICD was present in 9 patients (mean age 49.1 ± 13.7 years, 66.7% males), 100% with Boston Scientific third-generation EMBLEM MRI S-ICD, 11% with HeartMate II (HM II), 44% with HeartMate 3 (HM 3), and 44% with HeartWare (HW) LVAD. The incidence of noise from LVAD-related electromagnetic interference (EMI) was 33% and was only seen with HM 3 LVAD. Multiple measures attempted to resolve noise, including using alternative S-ICD sensing vector, adjusting S-ICD time zone, and increasing LVAD pump speed, were unsuccessful, necessitating S-ICD device therapies to be turned off permanently.
CONCLUSIONS: The incidence of LVAD-related S-ICD noise is high in patients with concomitant LVAD and S-ICD with significant impact on device function. As conservative management failed to resolve the EMI, the S-ICDs had to be programmed off to avoid inappropriate shocks. This study highlights the importance of awareness of LVAD-SICD device interference and the need to improve S-ICD detection algorithms to eliminate noise.
Full text links
Related Resources
Trending Papers
Renin-Angiotensin-Aldosterone System: From History to Practice of a Secular Topic.International Journal of Molecular Sciences 2024 April 5
Prevention and treatment of ischaemic and haemorrhagic stroke in people with diabetes mellitus: a focus on glucose control and comorbidities.Diabetologia 2024 April 17
British Society for Rheumatology guideline on management of adult and juvenile onset Sjögren disease.Rheumatology 2024 April 17
Albumin: a comprehensive review and practical guideline for clinical use.European Journal of Clinical Pharmacology 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app