Incidence and Predictors of Cardiac Implantable Electronic Devices Malfunction with Radiotherapy Treatment.
Journal of Clinical Medicine 2022 October 28
AIMS: To investigate the incidence of cardiac implantable electronic devices (CIED) malfunction with radiotherapy (RT) treatment and assess predictors of CIED malfunction.
METHODS: A 6-year retrospective analysis of patients who underwent RT with CIED identified through the radiation oncology electronic database. Clinical, RT (cumulative dose, dose per fraction, beam energy, beam energy dose, and anatomical area of RT) and CIED details (type, manufacturer, and device malfunction) were collected from electronic medical records.
RESULTS: We identified 441 patients with CIED who underwent RT. CIED encountered a permanent pacemaker (PPM) (78%), cardiac resynchronization therapy-pacing (CRT-P) (2%), an implantable cardioverter defibrillator (ICD) (10%), and a CRT-defibrillator (CRT-D) (10%). The mean cumulative dose of RT was 36 gray (Gy) (IQR 1.8-80 Gy) and the most common beam energy used was photon ≥10 megavolt (MV) with a median dose of 7 MV (IQR 5-18 MV). We further identified 17 patients who had CIED malfunction with RT. This group had a higher cumulative RT dose of 42.5 Gy (20-77 Gy) and a photon dose of 14 MV (12-18 MV). None of the malfunctions resulted in clinical symptoms. Using logistic regression, the predictors of CIED malfunction were photon beam energy use ≥10 MV (OR 5.73; 95% CI, 1.58-10.76), anatomical location of RT above the diaphragm (OR 5.2, 95% CI, 1.82-15.2), and having a CIED from the ICD group (OR 4.6, 95% CI, 0.75-10.2).
CONCLUSION: Clinicians should be aware of predictors of CIED malfunction with RT to ensure the safety of patients.
METHODS: A 6-year retrospective analysis of patients who underwent RT with CIED identified through the radiation oncology electronic database. Clinical, RT (cumulative dose, dose per fraction, beam energy, beam energy dose, and anatomical area of RT) and CIED details (type, manufacturer, and device malfunction) were collected from electronic medical records.
RESULTS: We identified 441 patients with CIED who underwent RT. CIED encountered a permanent pacemaker (PPM) (78%), cardiac resynchronization therapy-pacing (CRT-P) (2%), an implantable cardioverter defibrillator (ICD) (10%), and a CRT-defibrillator (CRT-D) (10%). The mean cumulative dose of RT was 36 gray (Gy) (IQR 1.8-80 Gy) and the most common beam energy used was photon ≥10 megavolt (MV) with a median dose of 7 MV (IQR 5-18 MV). We further identified 17 patients who had CIED malfunction with RT. This group had a higher cumulative RT dose of 42.5 Gy (20-77 Gy) and a photon dose of 14 MV (12-18 MV). None of the malfunctions resulted in clinical symptoms. Using logistic regression, the predictors of CIED malfunction were photon beam energy use ≥10 MV (OR 5.73; 95% CI, 1.58-10.76), anatomical location of RT above the diaphragm (OR 5.2, 95% CI, 1.82-15.2), and having a CIED from the ICD group (OR 4.6, 95% CI, 0.75-10.2).
CONCLUSION: Clinicians should be aware of predictors of CIED malfunction with RT to ensure the safety of patients.
Full text links
Related Resources
Trending Papers
Mineralocorticoid receptor antagonists and reno-protection: What's the evidence & where do they fit? A guide for non-specialists.Diabetes, Obesity & Metabolism 2024 May 8
Angiotensin Receptor Blocker-Neprilysin Inhibitor for Heart Failure with Reduced Ejection Fraction.Pharmacological Research : the Official Journal of the Italian Pharmacological Society 2024 May 12
The Therapy and Management of Heart Failure with Preserved Ejection Fraction: New Insights on Treatment.Cardiac Failure Review 2024
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app