We have located links that may give you full text access.
Trifecta versus Perimount Magna Ease aortic valves: Failure mechanisms.
Asian Cardiovascular & Thoracic Annals 2022 September
BACKGROUND: There are increasing reports of early externally mounted pericardial Trifecta bioprosthesis failure. We compared the hemodynamic performance of Trifecta and Carpentier-Edwards Perimount Magna Ease valves to determine the failure mechanism.
METHODS: We retrospectively included 270 consecutive patients (age: 73.4 ± 8.2 years; 57.5% male; mean follow-up: 48.0 ± 20.3 months) who underwent aortic valve replacement from 2014 to 2021 at a single center and compared the Trifecta ( N = 137) and Carpentier-Edwards Perimount Magna Ease valve ( N = 133) patients.
RESULTS: The prosthetic valve major aortic regurgitation incidence was higher for the Trifecta than that for the Carpentier-Edwards Perimount Magna Ease valve (6.3% vs. 0%, P < 0.009). Among the Trifecta failures, 33% developed structural valve deterioration, but all requiring redo aortic valve replacement developed major prosthetic valve aortic regurgitation. Freedom at 5 years from redo aortic valve replacement due to structural valve deterioration was significantly lower for Trifecta (89.4% vs. 100%, P = 0.003). The reoperation hazards were determined for Trifecta (vs. Carpentier-Edwards Perimount Magna Ease): 11.6 (1.47-90.9; P = 0.02), prosthetic valve aortic regurgitation: 2.38 (1.70-3.32; P < 0.01), structural valve deterioration: 20.82 (4.08-106.2; P < 0.01), 5-year mean transprosthetic pressure gradient: 1.14 per 1-point increase (1.03-1.24; P = 0.007), and urgent surgery: 10.1 (2.59-39.0; P = 0.001). The Cox regression analysis identified that prosthetic valve aortic regurgitation solely contributed to redo aortic valve replacement (hazard ratio: 2.38; confidence intervals: 1.70-3.32).
CONCLUSIONS: Significantly, more early failures occurred with the Trifecta valve than the Carpentier-Edwards Perimount Magna Ease valve but the Trifecta showed reasonable mean transprosthetic pressure gradient over time. Prosthetic valve aortic regurgitation and calcific structural valve deterioration synergistically contributed to Trifecta valve failure alternatively.
METHODS: We retrospectively included 270 consecutive patients (age: 73.4 ± 8.2 years; 57.5% male; mean follow-up: 48.0 ± 20.3 months) who underwent aortic valve replacement from 2014 to 2021 at a single center and compared the Trifecta ( N = 137) and Carpentier-Edwards Perimount Magna Ease valve ( N = 133) patients.
RESULTS: The prosthetic valve major aortic regurgitation incidence was higher for the Trifecta than that for the Carpentier-Edwards Perimount Magna Ease valve (6.3% vs. 0%, P < 0.009). Among the Trifecta failures, 33% developed structural valve deterioration, but all requiring redo aortic valve replacement developed major prosthetic valve aortic regurgitation. Freedom at 5 years from redo aortic valve replacement due to structural valve deterioration was significantly lower for Trifecta (89.4% vs. 100%, P = 0.003). The reoperation hazards were determined for Trifecta (vs. Carpentier-Edwards Perimount Magna Ease): 11.6 (1.47-90.9; P = 0.02), prosthetic valve aortic regurgitation: 2.38 (1.70-3.32; P < 0.01), structural valve deterioration: 20.82 (4.08-106.2; P < 0.01), 5-year mean transprosthetic pressure gradient: 1.14 per 1-point increase (1.03-1.24; P = 0.007), and urgent surgery: 10.1 (2.59-39.0; P = 0.001). The Cox regression analysis identified that prosthetic valve aortic regurgitation solely contributed to redo aortic valve replacement (hazard ratio: 2.38; confidence intervals: 1.70-3.32).
CONCLUSIONS: Significantly, more early failures occurred with the Trifecta valve than the Carpentier-Edwards Perimount Magna Ease valve but the Trifecta showed reasonable mean transprosthetic pressure gradient over time. Prosthetic valve aortic regurgitation and calcific structural valve deterioration synergistically contributed to Trifecta valve failure alternatively.
Full text links
Related Resources
Trending Papers
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
British Society for Rheumatology guideline on management of adult and juvenile onset Sjögren disease.Rheumatology 2024 April 17
Albumin: a comprehensive review and practical guideline for clinical use.European Journal of Clinical Pharmacology 2024 April 13
Renin-Angiotensin-Aldosterone System: From History to Practice of a Secular Topic.International Journal of Molecular Sciences 2024 April 5
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app