INTRODUCTION: Sonographic technologies can estimate extravascular lung water (EVLW) in hemodialysis (HD) patients. This study investigated the suitability of a handheld scanner in contrast to a portable scanner for quantifying EVLW in hospitalized patients requiring HD.
METHODS: In this prospective study, 54 hospitalized HD patients were enrolled. Bedside lung ultrasound was performed within 30 min before and after dialysis using handheld (phased array transducer, 1.7-3.8 MHz) and portable (curved probe, 5-2 MHz) ultrasound devices. Eight lung zones were scanned for total B-lines number (TBLN). The maximum diameter of inferior vena cava (IVC) was measured. We performed Passing-Bablok regression, Deming regression, Bland-Altman, and logistic regression analysis.
RESULTS: The 2 devices did not differ in measuring TBLN and IVC (p > 0.05), showing a high correlation (r = 0.92 and r = 0.51, respectively). Passing-Bablok regression had a slope of 1.11 and an intercept of 0 for TBLN, and the slope of Deming regression was 1.02 within the CI bands of 0.94 and 1.11 in the full cohort. TBLN was logarithmically transformed for Bland-Altman analysis, showing a bias of 0.06 (TBLN = 1.2) between devices. The slope and intercept of the Deming regression in IVC measurements were 0.77 and 0.46, respectively; Bland-Altman plot showed a bias of -0.07. Compared with predialysis, TBLN significantly (p < 0.001) decreased after dialysis, while IVC was unchanged (p = 0.16). Univariate analysis showed that cardiovascular disease (odds ratio [OR] 8.94 [2.13-61.96], p = 0.002), smoking history (OR 5.75 [1.8-20.46], p = 0.003), and right pleural effusion (OR 5.0 [1.2-25.99], p = 0.03) were strong predictors of EVLW indicated by TBLN ≥ 4.
CONCLUSION: The lung and IVC findings obtained from handheld and portable ultrasound scanners are comparable and concordant. Cardiovascular disease and smoking history were strong predictors of EVLW. The use of TBLN to assess EVLW in hospitalized HD patients is feasible. Further studies are needed to determine if TBLN can help guide volume removal in HD patients.
METHODS: In this prospective study, 54 hospitalized HD patients were enrolled. Bedside lung ultrasound was performed within 30 min before and after dialysis using handheld (phased array transducer, 1.7-3.8 MHz) and portable (curved probe, 5-2 MHz) ultrasound devices. Eight lung zones were scanned for total B-lines number (TBLN). The maximum diameter of inferior vena cava (IVC) was measured. We performed Passing-Bablok regression, Deming regression, Bland-Altman, and logistic regression analysis.
RESULTS: The 2 devices did not differ in measuring TBLN and IVC (p > 0.05), showing a high correlation (r = 0.92 and r = 0.51, respectively). Passing-Bablok regression had a slope of 1.11 and an intercept of 0 for TBLN, and the slope of Deming regression was 1.02 within the CI bands of 0.94 and 1.11 in the full cohort. TBLN was logarithmically transformed for Bland-Altman analysis, showing a bias of 0.06 (TBLN = 1.2) between devices. The slope and intercept of the Deming regression in IVC measurements were 0.77 and 0.46, respectively; Bland-Altman plot showed a bias of -0.07. Compared with predialysis, TBLN significantly (p < 0.001) decreased after dialysis, while IVC was unchanged (p = 0.16). Univariate analysis showed that cardiovascular disease (odds ratio [OR] 8.94 [2.13-61.96], p = 0.002), smoking history (OR 5.75 [1.8-20.46], p = 0.003), and right pleural effusion (OR 5.0 [1.2-25.99], p = 0.03) were strong predictors of EVLW indicated by TBLN ≥ 4.
CONCLUSION: The lung and IVC findings obtained from handheld and portable ultrasound scanners are comparable and concordant. Cardiovascular disease and smoking history were strong predictors of EVLW. The use of TBLN to assess EVLW in hospitalized HD patients is feasible. Further studies are needed to determine if TBLN can help guide volume removal in HD patients.
Full text links
Related Resources
Trending Papers
Renin-Angiotensin-Aldosterone System: From History to Practice of a Secular Topic.International Journal of Molecular Sciences 2024 April 5
Prevention and treatment of ischaemic and haemorrhagic stroke in people with diabetes mellitus: a focus on glucose control and comorbidities.Diabetologia 2024 April 17
British Society for Rheumatology guideline on management of adult and juvenile onset Sjögren disease.Rheumatology 2024 April 17
Albumin: a comprehensive review and practical guideline for clinical use.European Journal of Clinical Pharmacology 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app