Silicone oil-free syringes, siliconized syringes and needles: quantitative assessment of silicone oil release with drugs used for intravitreal injection.
Acta Ophthalmologica 2021 April 23
PURPOSE: This study aimed to quantify the amount of silicone oil (SO) released across a variety of syringe and needle models routinely used for intravitreal injection.
METHODS: The release of SO was assessed in eight models of syringes, two of which were reported to be 'SO-free', and eleven models of needles with unknown SO content. To evaluate SO release within the context of anti-VEGF therapeutics, syringes were evaluated using aflibercept, bevacizumab, buffer, ziv-aflibercept and formulation buffer. All syringe tests were performed with or without agitation by flicking for syringes. Needles were evaluated without agitation only. Samples were fluorescently labelled to identify SO, and triplicate measurements were collected using imaging flow cytometry.
RESULTS: Seven out of 8 syringe models showed a statistically significant increase in the SO particle count after agitation. The two SO-free syringe models (HSW Norm-Ject, Daikyo Crystal Zenith) released the least SO particles, with or without agitation, whereas the BD Ultra-Fine and Saldanha-Rodrigues syringes released the most. More SO was released when the syringes were prefilled with formulation buffer than with ziv-aflibercept. Syringes filled with aflibercept and bevacizumab had intermediate levels. Agitation increased the release of SO into each of the drug solutions. Silicone oil (SO) was detected in all needles.
CONCLUSIONS: Agitation of the syringe by flicking leads to a substantial increase in the number of SO particles. Silicone oil (SO)-free syringes had the best performance, but physicians must also be aware that needles are siliconized and also contribute to the injection of SO into the vitreous.
METHODS: The release of SO was assessed in eight models of syringes, two of which were reported to be 'SO-free', and eleven models of needles with unknown SO content. To evaluate SO release within the context of anti-VEGF therapeutics, syringes were evaluated using aflibercept, bevacizumab, buffer, ziv-aflibercept and formulation buffer. All syringe tests were performed with or without agitation by flicking for syringes. Needles were evaluated without agitation only. Samples were fluorescently labelled to identify SO, and triplicate measurements were collected using imaging flow cytometry.
RESULTS: Seven out of 8 syringe models showed a statistically significant increase in the SO particle count after agitation. The two SO-free syringe models (HSW Norm-Ject, Daikyo Crystal Zenith) released the least SO particles, with or without agitation, whereas the BD Ultra-Fine and Saldanha-Rodrigues syringes released the most. More SO was released when the syringes were prefilled with formulation buffer than with ziv-aflibercept. Syringes filled with aflibercept and bevacizumab had intermediate levels. Agitation increased the release of SO into each of the drug solutions. Silicone oil (SO) was detected in all needles.
CONCLUSIONS: Agitation of the syringe by flicking leads to a substantial increase in the number of SO particles. Silicone oil (SO)-free syringes had the best performance, but physicians must also be aware that needles are siliconized and also contribute to the injection of SO into the vitreous.
Full text links
Related Resources
Trending Papers
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
British Society for Rheumatology guideline on management of adult and juvenile onset Sjögren disease.Rheumatology 2024 April 17
Albumin: a comprehensive review and practical guideline for clinical use.European Journal of Clinical Pharmacology 2024 April 13
Renin-Angiotensin-Aldosterone System: From History to Practice of a Secular Topic.International Journal of Molecular Sciences 2024 April 5
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app