Accuracy of patient-specific instrumentation in total ankle arthroplasty: A comparative study.
Foot and Ankle Surgery : Official Journal of the European Society of Foot and Ankle Surgeons 2018 Februrary 17
BACKGROUND: Patient-specific instrumentation (PSI) for TAA is a novel technology with several potential benefits. The primary goal of this study was to compare the use of PSI with the standard referencing guide (SRG) in regards to accuracy of tibial implant positioning. Operative time, fluoroscopy time and accuracy of PSI preoperative reports were also evaluated.
METHODS: A retrospective analysis of 99 patients who underwent a primary TAA with the INFINITY prosthesis (Wright Medical, Memphis, TN) was performed. Patients were divided in two groups based on the type of instrumentation used during the TAA (75 in the PSI group vs 24 in the SRG group). There was no significant difference between groups in regards to age at the time of surgery (P=0.122), sex (P=0.138), number of concomitant procedures performed during surgery (P=0.567) and etiology (P=0.841). However, preoperative deformity was significantly smaller in the PSI group (P=0.002).
RESULTS: Tibial implant positioning was similar between groups. In the coronal plane, the absolute deviation of the tibial implant from the intended alignment was 1.7±1.4° for the SRG and 1.6±1.2° for PSI (P=0.710). In the sagittal plane, the absolute alignment deviation of the tibial implant was 1.8±1.4° for the SRG and 1.9±1.5° for PSI (P=0.675). Operative time (167 vs 190min, P=0.040) and fluoroscopy time (85 vs 158s, P<0.001) were significantly decreased in the PSI group. The PSI preoperative plan report correctly predicted the implant size in 73% of cases for the tibial component and in 51% of cases for the talar component.
CONCLUSIONS: PSI provided similar tibial component alignment as standard instrumentation. Additionally, PSI preoperative plan reports were poor predictors of implant sizing. Therefore, the final decision should always be based on surgeon's experience in order to prevent errors in implant sizing and positioning.
LEVEL OF EVIDENCE: Level III, retrospective comparative study.
METHODS: A retrospective analysis of 99 patients who underwent a primary TAA with the INFINITY prosthesis (Wright Medical, Memphis, TN) was performed. Patients were divided in two groups based on the type of instrumentation used during the TAA (75 in the PSI group vs 24 in the SRG group). There was no significant difference between groups in regards to age at the time of surgery (P=0.122), sex (P=0.138), number of concomitant procedures performed during surgery (P=0.567) and etiology (P=0.841). However, preoperative deformity was significantly smaller in the PSI group (P=0.002).
RESULTS: Tibial implant positioning was similar between groups. In the coronal plane, the absolute deviation of the tibial implant from the intended alignment was 1.7±1.4° for the SRG and 1.6±1.2° for PSI (P=0.710). In the sagittal plane, the absolute alignment deviation of the tibial implant was 1.8±1.4° for the SRG and 1.9±1.5° for PSI (P=0.675). Operative time (167 vs 190min, P=0.040) and fluoroscopy time (85 vs 158s, P<0.001) were significantly decreased in the PSI group. The PSI preoperative plan report correctly predicted the implant size in 73% of cases for the tibial component and in 51% of cases for the talar component.
CONCLUSIONS: PSI provided similar tibial component alignment as standard instrumentation. Additionally, PSI preoperative plan reports were poor predictors of implant sizing. Therefore, the final decision should always be based on surgeon's experience in order to prevent errors in implant sizing and positioning.
LEVEL OF EVIDENCE: Level III, retrospective comparative study.
Full text links
Related Resources
Trending Papers
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Review article: Recent advances in ascites and acute kidney injury management in cirrhosis.Alimentary Pharmacology & Therapeutics 2024 March 26
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app