Diagnostic evaluation of an in-house developed single-tube, duplex, nested IS6110 real-time PCR assay for rapid pulmonary tuberculosis diagnosis.
Tuberculosis 2018 September
OBJECTIVE: To perform a prospective evaluation on the diagnostic performance of an in-house developed, duplex nested IS6110 real-time Polymerase-Chain-Reaction (PCR) assay (IS6110-qPCR assay) for rapid pulmonary TB diagnosis.
METHODS: A total of 503 sputum specimens were prospectively collected from July 2016 to November 2016. Diagnostic accuracy and optimal cut-off Cycle-threshold (Ct) value for IS6110-qPCR assay was determined by Receiver Operating Characteristic (ROC) curve. Using the optimal cut-off Ct, diagnostic performance of IS6110-qPCR assay was assessed with reference to both bacteriological and clinical information. Meanwhile, limit of detection (LOD) was calculated using Mycobacterium tuberculosis H37Rv as reference strain.
RESULT: ROC curve analysis of IS6110-qPCR assay showed a high Area Under the Curve (AUC) value (0.948) with optimal Ct value at 24.140. Prospective analysis of IS6110-qPCR assay with cut-off Ct = 24.140 showed a high overall sensitivity and specificity of 97.2% and 99.7%, respectively. No cross reactivity was observed among all non-tuberculous mycobacteria specimens in this study. LOD analysis on MTB-spiked sputum showed an average detection limit of 5.0 CFU/mL at Ct = 23.18 (±SD, 0.57).
CONCLUSION: IS6110-qPCR assay is a highly accurate and cost-effective assay developed for primary screening of suspected TB cases, which is particularly suitable for regions with limited resources but high TB burden.
METHODS: A total of 503 sputum specimens were prospectively collected from July 2016 to November 2016. Diagnostic accuracy and optimal cut-off Cycle-threshold (Ct) value for IS6110-qPCR assay was determined by Receiver Operating Characteristic (ROC) curve. Using the optimal cut-off Ct, diagnostic performance of IS6110-qPCR assay was assessed with reference to both bacteriological and clinical information. Meanwhile, limit of detection (LOD) was calculated using Mycobacterium tuberculosis H37Rv as reference strain.
RESULT: ROC curve analysis of IS6110-qPCR assay showed a high Area Under the Curve (AUC) value (0.948) with optimal Ct value at 24.140. Prospective analysis of IS6110-qPCR assay with cut-off Ct = 24.140 showed a high overall sensitivity and specificity of 97.2% and 99.7%, respectively. No cross reactivity was observed among all non-tuberculous mycobacteria specimens in this study. LOD analysis on MTB-spiked sputum showed an average detection limit of 5.0 CFU/mL at Ct = 23.18 (±SD, 0.57).
CONCLUSION: IS6110-qPCR assay is a highly accurate and cost-effective assay developed for primary screening of suspected TB cases, which is particularly suitable for regions with limited resources but high TB burden.
Full text links
Related Resources
Trending Papers
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app