Effect of Dual-Source Radiofrequency Transmission on Left Ventricular Measurements and Measurement Reproducibility at 3.0 T Cardiac MR Imaging: Comparison with Conventional Single-Source Transmission Reference.

Academic Radiology 2018 August 30
RATIONALE AND OBJECTIVES: To prospectively assess effect of dual-source radiofrequency (RF) transmission on left ventricular (LV) measurements and measurements reproducibility at 3.0 T MR using balanced steady-state free precession (b-SSFP) cine imaging, compared to the conventional single-source RF transmission reference approach.

MATERIALS AND METHODS: Cardiac b-SSFP cine imaging was performed in 19 subjects at 3.0 T MR equipped with dual-source RF transmission. All images were analyzed to obtain LV end-diastolic volume, end-systolic volume, stroke volume, ejection fraction, mass, LV end-diastolic inferior wall thickness, and interventricular septal thickness. The difference of all LV measurements between the two imaging techniques was tested with the paired t test and the intertechnique agreement was tested through linear regression and Bland-Altman analyses. Additionally, repeated LV measurements were performed to determine intra and interobserver variability with the Bland-Altman method, the 95% limits of agreement, the coefficient of variation (CV) and the intraclass correlation coefficient.

RESULTS: Compared to conventional single-source, dual-source slightly overestimated end-diastolic volume, end-systolic volume, and stroke volume (mean differences, 3.9 mL ± 9.7, 1.1 mL ± 2.6, and 2.8 mL ± 9.1, respectively; p > 0.05), resulting in a small but significant positive bias in ejection fraction (1.5% ± 2.6; p = 0.021). Mass was significantly smaller with dual-source than with single-source (-4.0 g ± 6.5, p = 0.001). Dual-source slightly underestimated interventricular septal thickness (-0.29 mm ± 0.6, p = 0.067) and significantly underestimated LV end-diastolic inferior wall thickness (-0.55 mm ± 0.4, p < 0.0001). The two techniques in measurements correlated highly (r2  = 0.81 to 0.96, p < 0.0001). Intra and interobserver variability in dual-source measurements was much lower than that in single-source, and variability values were <14.0%.

CONCLUSION: Improved image quality of b-SSFP cine imaging at 3.0 T MR with dual-source RF transmission may provide more reproducible LV measurements compared to conventional single-source approach. Dual-source RF transmission also provides a reasonable estimate of the LV measurements.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app