Analysis of the long-term stability of the output of electron beams generated by the Novac11™ IORT accelerator.
Reports of Practical Oncology and Radiotherapy 2018 September
Aim: The aim of the study was to analyze the long-term stability of electron beams generated by the Novac11™ IORT accelerator.
Background: Novac11™ (NRT® ) is a mobile electron accelerator designed to irradiate small areas of tissue, up to 10 cm in diameter, with electron beams during surgical procedures. It is characterized by a great mobility guaranteed by a number of degrees of freedom enabling irradiation in the conditions of an operating theatre.
Materials and methods: Over the period of January 2013 and September 2016, the measurement sessions of the output of clinically used beam qualities (6, 8 and 10 MeV) were carried out 41 times. Because of the unsatisfactory long-term stability, an extra procedure of tuning of the magnetron, suggested by the manufacturer, was introduced in October 2015, 15 measurements were performed since then. The output of the Novac11™ accelerator was measured in the reference conditions recommended by the IAEA Report 398, the measurements of the charge in the ionization chamber at the reference depth were carried out with a Dose1™ electrometer and a plane-parallel chamber PPC05™ from IBA® .
Results: The introduction of the tuning of the magnetron procedure resulted in satisfactory long-term stability of the measured outputs below 2%.
Conclusions: After the introduction of the STV parameter tuning procedure, the long-term stability of the Novac11™ output increased considerably and is within the values declared by the manufacturer.
Background: Novac11™ (NRT® ) is a mobile electron accelerator designed to irradiate small areas of tissue, up to 10 cm in diameter, with electron beams during surgical procedures. It is characterized by a great mobility guaranteed by a number of degrees of freedom enabling irradiation in the conditions of an operating theatre.
Materials and methods: Over the period of January 2013 and September 2016, the measurement sessions of the output of clinically used beam qualities (6, 8 and 10 MeV) were carried out 41 times. Because of the unsatisfactory long-term stability, an extra procedure of tuning of the magnetron, suggested by the manufacturer, was introduced in October 2015, 15 measurements were performed since then. The output of the Novac11™ accelerator was measured in the reference conditions recommended by the IAEA Report 398, the measurements of the charge in the ionization chamber at the reference depth were carried out with a Dose1™ electrometer and a plane-parallel chamber PPC05™ from IBA® .
Results: The introduction of the tuning of the magnetron procedure resulted in satisfactory long-term stability of the measured outputs below 2%.
Conclusions: After the introduction of the STV parameter tuning procedure, the long-term stability of the Novac11™ output increased considerably and is within the values declared by the manufacturer.
Full text links
Related Resources
Trending Papers
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Review article: Recent advances in ascites and acute kidney injury management in cirrhosis.Alimentary Pharmacology & Therapeutics 2024 March 26
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app