Pencil beam scanning proton therapy for mediastinal lymphomas in deep inspiration breath-hold: a retrospective assessment of plan robustness.
Acta Oncologica 2024 Februrary 29
PURPOSE/BACKGROUND: The aim of this study was to evaluate pencil beam scanning (PBS) proton therapy (PT) in deep inspiration breath-hold (DIBH) for mediastinal lymphoma patients, by retrospectively evaluating plan robustness to the clinical target volume (CTV) and organs at risk (OARs) on repeated CT images acquired throughout treatment. Methods: Sixteen mediastinal lymphoma patients treated with PBS-PT in DIBH were included. Treatment plans (TPs) were robustly optimized on the CTV (7 mm/4.5%). Repeated verification CTs (vCT) were acquired during the treatment course, resulting in 52 images for the entire patient cohort. The CTV and OARs were transferred from the planning CT to the vCTs with deformable image registration and the TPs were recalculated on the vCTs. Target coverage and OAR doses at the vCTs were compared to the nominal plan. Deviation in lung volume was also calculated.
RESULTS: The TPs demonstrated high robust target coverage throughout treatment with D98%,CTV deviations within 2% for 14 patients and above the desired requirement of 95% for 49/52 vCTs. However, two patients did not achieve a robust dose to CTV due to poor DIBH reproducibility, with D98%,CTV at 78 and 93% respectively, and replanning was performed for one patient. Adequate OAR sparing was achieved for all patients. Total lung volume variation was below 10% for 39/52 vCTs.
CONCLUSION: PBS PT in DIBH is generally a robust technique for treatment of mediastinal lymphomas. However, closely monitoring the DIBH-reproducibility during treatment is important to avoid underdosing CTV and achieve sufficient dose-sparing of the OARs.
RESULTS: The TPs demonstrated high robust target coverage throughout treatment with D98%,CTV deviations within 2% for 14 patients and above the desired requirement of 95% for 49/52 vCTs. However, two patients did not achieve a robust dose to CTV due to poor DIBH reproducibility, with D98%,CTV at 78 and 93% respectively, and replanning was performed for one patient. Adequate OAR sparing was achieved for all patients. Total lung volume variation was below 10% for 39/52 vCTs.
CONCLUSION: PBS PT in DIBH is generally a robust technique for treatment of mediastinal lymphomas. However, closely monitoring the DIBH-reproducibility during treatment is important to avoid underdosing CTV and achieve sufficient dose-sparing of the OARs.
Full text links
Related Resources
Trending Papers
Haemodynamic monitoring during noncardiac surgery: past, present, and future.Journal of Clinical Monitoring and Computing 2024 April 31
2024 AHA/ACC/AMSSM/HRS/PACES/SCMR Guideline for the Management of Hypertrophic Cardiomyopathy: A Report of the American Heart Association/American College of Cardiology Joint Committee on Clinical Practice Guidelines.Circulation 2024 May 9
Obesity pharmacotherapy in older adults: a narrative review of evidence.International Journal of Obesity 2024 May 7
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app