Add like
Add dislike
Add to saved papers

An Extremely Rare Case of Cementless Third Generation Corail Stem Neck Fracture With Fractographic Analysis.

Arthroplasty Today 2023 August
The cementless Corail stem is one of the most frequently implanted stems and has undergone several design changes. Currently in use is the third generation, named Corail AMT. Until now, only one third-generation Corail stem neck fracture has been described in 2020. In our paper, we present an almost identical complication with an additional analysis of the fracture using a scanning electron microscope. The revision surgery consisted of changing the broken implant with a Corail revision stem, along with replacing the polyethylene liner and the femoral head with new one, after which the patient achieved a full recovery. According to the available literature, this is the second case of this extremely rare complication.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app