Early detection of oliguric events in critically ill patients in the ICU with a novel continuous urine flow measurement device: results of an initial validation study.
Journal of Clinical Monitoring and Computing 2023 April 8
OBJECTIVES: Urine output is used to evaluate fluid status and is an important marker for acute kidney injury (AKI). Our primary aim was to validate a new automatic urine output monitoring device by comparison to the current practice - the standard urometer.
METHODS: We conducted a prospective observational study in three ICUs. Urine flow measurements by Serenno Medical Automatic urine output measuring device (Serenno Medical, Yokneam, Israel) were compared to standard urometer readings taken automatically at 5-minutes intervals by a camera, and to hourly urometer readings by the nurses, both over 1 to 7 days. Our primary outcome was the difference between urine flow assessed by the Serenno device and reference camera-derived measurements (Camera). Our secondary outcome was the difference between urine flow assessed by the Serenno device and hourly nursing assessments (Nurse), and detection of oliguria.
RESULTS: Thirty-seven patients completed the study, with 1,306 h of recording and a median of 25 measurement hours per patient. Bland and Altman analysis comparing the study device to camera measurements demonstrated good agreement, with a bias of -0.4 ml/h and 95% confidence intervals ranging from - 28 to 27ml/h. Concordance was 92%. The correlation between Camera and hourly nursing assessment of urine output was distinctly worse with a bias of 7.2 ml and limits of agreement extending from - 75 to + 107 ml. Severe oliguria (urine output < 0.3 ml/kg/h) lasting 2 h or more was common and observed in 8 (21%) of patients. Among the severe oliguric events lasting more than 3 consecutive hours, 6 (41%) were not detected or documented by the nursing staff. There were no device-related complications.
CONCLUSION: The Serenno Medical Automatic urine output measuring device required minimal supervision, little ICU nursing staff attention, and is sufficiently accurate and precise. In addition to providing continuous assessments of urine output, it was considerably more accurate than hourly nursing assessments.
METHODS: We conducted a prospective observational study in three ICUs. Urine flow measurements by Serenno Medical Automatic urine output measuring device (Serenno Medical, Yokneam, Israel) were compared to standard urometer readings taken automatically at 5-minutes intervals by a camera, and to hourly urometer readings by the nurses, both over 1 to 7 days. Our primary outcome was the difference between urine flow assessed by the Serenno device and reference camera-derived measurements (Camera). Our secondary outcome was the difference between urine flow assessed by the Serenno device and hourly nursing assessments (Nurse), and detection of oliguria.
RESULTS: Thirty-seven patients completed the study, with 1,306 h of recording and a median of 25 measurement hours per patient. Bland and Altman analysis comparing the study device to camera measurements demonstrated good agreement, with a bias of -0.4 ml/h and 95% confidence intervals ranging from - 28 to 27ml/h. Concordance was 92%. The correlation between Camera and hourly nursing assessment of urine output was distinctly worse with a bias of 7.2 ml and limits of agreement extending from - 75 to + 107 ml. Severe oliguria (urine output < 0.3 ml/kg/h) lasting 2 h or more was common and observed in 8 (21%) of patients. Among the severe oliguric events lasting more than 3 consecutive hours, 6 (41%) were not detected or documented by the nursing staff. There were no device-related complications.
CONCLUSION: The Serenno Medical Automatic urine output measuring device required minimal supervision, little ICU nursing staff attention, and is sufficiently accurate and precise. In addition to providing continuous assessments of urine output, it was considerably more accurate than hourly nursing assessments.
Full text links
Related Resources
Trending Papers
Consensus Statement on Vitamin D Status Assessment and Supplementation: Whys, Whens, and Hows.Endocrine Reviews 2024 April 28
The Tricuspid Valve: A Review of Pathology, Imaging, and Current Treatment Options: A Scientific Statement From the American Heart Association.Circulation 2024 April 26
Intravenous infusion of dexmedetomidine during the surgery to prevent postoperative delirium and postoperative cognitive dysfunction undergoing non-cardiac surgery: a meta-analysis of randomized controlled trials.European Journal of Medical Research 2024 April 19
Interstitial Lung Disease: A Review.JAMA 2024 April 23
Management of Diverticulitis: A Review.JAMA Surgery 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app