The Relationship between Bacterial Load and Initial Run Time of a Surgical Helmet.
BACKGROUND: Periprosthetic joint infection (PJI) is a complication of arthroplasty surgery with significant morbidity and mortality. Surgical helmets are a possible source of infection. Pre-existing dust and microorganisms on its surface may be blown into the surgical field by the helmet ventilation system.
METHODS: Twenty surgical helmets at our institution were assessed through microscopy and polymerase chain reaction testing. Helmets were arranged with agar plates under the front and rear outflow vents. Helmets ran while plates were exchanged at different time points. Bacterial growth was assessed via colony counts and correlated with fan operating time. Gram staining and 16S sequencing were performed to identify bacterial species.
RESULTS: The primary microbiological contaminate identified was Burkholderia. There was an inverse relationship between colony formation and fan operating time. The highest number of colonies was found within the first minute of fan operating time. There was a significant decrease in the number of colonies formed from the zero-minute to the three (27 vs 5; P = <.01), four (27 vs 3; P = <.01), and five-minute (27 vs 4; P = <.01) time points for the front outflow plates. A significant difference was also observed between the one-minute and four-minute time points ( P = .046).
CONCLUSION: We observed an inverse relationship between bacterial spread helmet fan operation time, which may correlate with dispersion of pre-existing contaminates. To decrease contamination risk, we recommend that helmets are run for at least 3 min prior to entering the operating room.
METHODS: Twenty surgical helmets at our institution were assessed through microscopy and polymerase chain reaction testing. Helmets were arranged with agar plates under the front and rear outflow vents. Helmets ran while plates were exchanged at different time points. Bacterial growth was assessed via colony counts and correlated with fan operating time. Gram staining and 16S sequencing were performed to identify bacterial species.
RESULTS: The primary microbiological contaminate identified was Burkholderia. There was an inverse relationship between colony formation and fan operating time. The highest number of colonies was found within the first minute of fan operating time. There was a significant decrease in the number of colonies formed from the zero-minute to the three (27 vs 5; P = <.01), four (27 vs 3; P = <.01), and five-minute (27 vs 4; P = <.01) time points for the front outflow plates. A significant difference was also observed between the one-minute and four-minute time points ( P = .046).
CONCLUSION: We observed an inverse relationship between bacterial spread helmet fan operation time, which may correlate with dispersion of pre-existing contaminates. To decrease contamination risk, we recommend that helmets are run for at least 3 min prior to entering the operating room.
Full text links
Related Resources
Trending Papers
Obesity pharmacotherapy in older adults: a narrative review of evidence.International Journal of Obesity 2024 May 7
SGLT2 Inhibitors in Kidney Diseases-A Narrative Review.International Journal of Molecular Sciences 2024 May 2
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app