Add like
Add dislike
Add to saved papers

Measurement of the anchoring force of covered self-expandable and lumen-apposing metal stents for interventional endoscopic ultrasonography.

OBJECTIVES: Interventional endoscopic ultrasonography is performed for various situations that require drainage, after which stent migration is the most severe adverse event. Several lumen-apposing metal stents (LAMS) and covered self-expandable metal stents (CSEMS) provide antimigration systems; however, their anchoring ability has not been studied well. Therefore, we measured and compared the anchoring force (ACF) of commercially available LAMS and CSEMS.

METHODS: Anchoring force was measured for five types of LAMS (NAGI, SPAXUS, Plumber, and AXIOS 8 and 10 mm) and seven types of CSEMS (BCL, SHCL, BCG, BPD [four types of HANAROSTENT], HILZO, Niti-S [Spring Stopper], and Wallflex). We created a phantom model for inducing stent migration. It has a rotatable part as a curved fixture, and we measured ACF at angles between 0° and 40°.

RESULTS: The mean ACF at 0° and 20° were NAGI 1.50, 1.84 N, SPAXUS 1.73, 1.72 N, Plumber 2.64, 2.03 N, and AXIOS 3.96, 3.61 N, respectively; and BCL 0.48, 0.53 N, Wallflex 0.53, 0.48 N, SHCL 0.64, 0.73 N, HILZO 1.09, 1.09 N, BCG 1.22, 1.20 N, BPD 1.78, 1.67 N, and Spring Stopper 2.29, 2.51 N.

CONCLUSION: We measured ACF in LAMS and CSEMS with a new phantom model. The highest value among the LAMS was for the AXIOS and among the CSEMS was for the Spring Stopper. Some of the ACF values varied with the direction of pull. These findings may have a significant impact on stent selection for interventional endoscopic ultrasonography, and this model can be used to evaluate newly developed stents.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app