Leadless pacemaker extraction from a single-center perspective.
Pacing and Clinical Electrophysiology : PACE 2018 Februrary
BACKGROUND: Leadless pacemaker can be considered as a technical revolution in cardiac pacing devices, with clear advantages over conventional pacemakers in overcoming all lead-related complications. However, the management of these devices once they reach the end of life (EOL) of the battery is still controversial. In the next years, there will be an increase in the need to define a clear strategy in the management of leadless PM once they reach their EOL. Safe extraction of these devices will define in a great manner this strategy METHODS: We performed the extraction of three functioning Nanostim leadless pacemaker prophylactically in two females and one male patients as part of the Nanostim battery depletion field action recommendation. All patients had a prior transesophageal 3D echocardiography to determine the device intracardiac mobility and the extent of possible endothelialization. For the extractions, we used the Nanostim Retrieval Catheter S1RSIN (St. Jude Medical, St. Paul, MN, USA), which is a proprietary catheter provided by the manufacturing company based on a lasso.
RESULTS: Complete extraction of the devices was achieved in all patients using a relatively short fluoroscopic time (16, 19, and 12 minutes).
CONCLUSIONS: The extraction of leadless pacemakers can be considered as a safe and feasible procedure using the tools provided by the manufacturer and designed for the extraction. However, a very low threshold must be maintained to avoid any risk to the patients. Our extraction time ranges are between 983 and 1,070 days, nevertheless it is necessary to gather more long-term data to assess the feasibility and safety of these procedures.
RESULTS: Complete extraction of the devices was achieved in all patients using a relatively short fluoroscopic time (16, 19, and 12 minutes).
CONCLUSIONS: The extraction of leadless pacemakers can be considered as a safe and feasible procedure using the tools provided by the manufacturer and designed for the extraction. However, a very low threshold must be maintained to avoid any risk to the patients. Our extraction time ranges are between 983 and 1,070 days, nevertheless it is necessary to gather more long-term data to assess the feasibility and safety of these procedures.
Full text links
Related Resources
Trending Papers
Autoimmune Hemolytic Anemias: Classifications, Pathophysiology, Diagnoses and Management.International Journal of Molecular Sciences 2024 April 13
Executive Summary: State-of-the-Art Review: Unintended Consequences: Risk of Opportunistic Infections Associated with Long-term Glucocorticoid Therapies in Adults.Clinical Infectious Diseases 2024 April 11
Clinical practice guidelines on the management of status epilepticus in adults: A systematic review.Epilepsia 2024 April 13
Finerenone: From the Mechanism of Action to Clinical Use in Kidney Disease.Pharmaceuticals 2024 March 27
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app