[Application of injection test in confirming the ideal position of esophageal balloon catheter].

OBJECTIVE: To evaluate the safety and feasibility of injection test which is used to locate esophageal balloon catheter.

METHODS: A prospective study was conducted. The patients undergoing invasive mechanical ventilation (MV) admitted to general intensive care unit (ICU) of Beijing Tiantan Hospital Affiliated to Capital Medical University from May 2015 and March 2017 were enrolled. The commercially available esophageal balloon catheter was modified to perform injection test. The catheter was withdrawn step by step and the injection test was repeated until the presence disturbance wave presented, which indicated that the balloon had just entered the esophagus. The position where disturbance wave appears was named 0 cm. End-expiratory occlusions were performed at the positions of +15, +10, +5, 0, -5, -10 and -15 cm, respectively, and the changes of esophageal pressure (Pes) and airway pressures (Paw) were measured in the spontaneous breathing and passive ventilation, and the ratio between the changes (ΔPes/ΔPaw) was calculated.

RESULTS: A total of 20 patients were enrolled, of which 15 patients finished both the spontaneous and the passive ventilation parts, and 2 patients finished only the spontaneous part and 3 patients finished only passive part. (1) Disturbance waves could be induced by injection test in all patients. The average depth of disturbance wave in spontaneous breathing was deeper than that in passive ventilation (cm: 42.4±3.8 vs. 41.8±3.3), but there was no significant difference between the two ventilation settings (P = 0.132). No adverse events occurred during the study period. (2) Pes increased with the stepwise withdraw of esophageal catheter, reached the maximal value at +5 cm, and then decreased when the catheter was further withdrawn, no matter in the spontaneous or the passive ventilation. In spontaneous breathing, the ΔPes/ΔPaw was within the ideal range (0.8-1.2) at the positions of 0, -5 and -10 cm. The ΔPes/ΔPaw was closest to unity at the positions of 0 cm (0.98±0.15). The ΔPes/ΔPaw at -15 cm (0.66±0.26) was significantly lower than that at 0 cm (P < 0.05). For passive ventilation, the ΔPes/ΔPaw was within the ideal range at the positions of -5 cm and -10 cm, and the ΔPes/ΔPaw was closest to unity at the positions of -10 cm (0.94±0.12). The ΔPes/ΔPaw at 0 cm and -5 cm was significantly higher than that at -10 cm (1.43±0.31 and 1.12±0.14, respectively); while the ΔPes/ΔPaw at -15 cm (0.68±0.23) was significantly lower than that at -10 cm (all P < 0.01).

CONCLUSIONS: Ideal position of the esophageal balloon catheter could be determined quickly and easily by using injection test. The method is safe and clinically feasible.

Clinical Trial Registration: Clinical Trials, NCT02446938.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app