Stabilizing properties of the halo apparatus.

Spine 1997 April 2
STUDY DESIGN: A cadaveric cervical spine specimen fixed between a fiberglass torso and a plastic skull was used as a model to determine the effect of halo structural parameters on motion at a lesion simulated at C5-C6. In a second part, nine commercially available halo devices were compared.

OBJECTIVES: To define the contributions of the various components of the halo apparatus to reducing motion in an injured cervical spine and to compare the stability offered by a sample of commercially available halo devices. Controversy exists concerning the ability of the halo apparatus to stabilize the injured cervical spine.

SUMMARY OF BACKGROUND DATA: The halo apparatus has been shown to be the most effective nonsurgical method for stabilizing the fractured spine. Nonetheless, several clinical studies have demonstrated that unacceptably large motions can occur at the injured spinal segment stabilized with a halo apparatus.

METHODS: Each cadaveric cervical spine was mounted onto a fiberglass torso and a rigid plastic skull was attached to the base of the occiput. A posterior ligamentous lesion was created between C5 and C6. The halo ring was fitted to the skull and a vest to the torso. Loads were applied to the skull in flexion, extension, and lateral bending, and relative angulation between C5 and C6 was measured with electroinclinometers. In the first part, the effect of parameters such as vest tightness, vest-thorax friction, vest deformation, and connecting bar rigidity on spinal angulation were measured using one vest. In the second part, the stability offered by each of nine commercially available halo devices was compared.

RESULTS: Increasing chest strap tightness and decreasing vest deformation reduced angulation at the spinal lesion. Once connecting bar joints were tightened to 25% of their recommended torque, increased tightening or adding additional bars had no effect on rigidity. Although specific vests permitted significantly greater motion in specific directions, no vest allowed greater angulation consistently in all loading planes.

CONCLUSIONS: Increasing vest tightness, decreasing the deformability of the vest, and ensuring a good fit can reduce motion in the fractured spine. Most commercially available halo vests provide similar mechanical stability to the injured cervical spine.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app