Improved Reliability of Electromyography-Based Neuromuscular Monitoring During Laparoscopic Surgery Achieved by the Modified Attachment Method for Nihon-Kohden NM-345Y™ Stimulating Electrodes: A Case Report.

Curēus 2024 Februrary
Neuromuscular monitoring is crucial during the administration of neuromuscular blocking agents owing to individual variations in their effects. In electromyography (EMG)-based neuromuscular monitoring using the EMG electrodes (NM-345Y™, Nihon-Kohden Corporation, Tokyo, Japan) following the manufacturer-recommended attachment method, the accuracy of neuromuscular monitoring may be reduced when forearm limb position is changed. We previously devised a novel attachment method for NM-345Y™ stimulating electrodes in adult volunteers to maintain stable monitoring accuracy despite changes in forearm position. Its effectiveness in clinical practice was evaluated by conducting a descriptive study on a 52-year-old woman undergoing laparoscopic uterine surgery. NM-345Y™ electrodes were attached to each forearm following the manufacturer's recommendations (Pattern N) and our novel method (Pattern C). In Pattern C, NM-345Y™ was attached without ultrasound guidance so that the ulnar nerve crossed the line connecting the centers of the anode and cathode of the stimulating electrode. Pattern C exhibited consistent EMG-based monitoring accuracy even with changes in forearm position despite a smaller stimulus current value at calibration. Additionally, Pattern C displayed reliable recovery of the train-of-four (TOF) response after sugammadex administration in the original forearm position, with no observed adverse events. In contrast, Pattern N showed unstable monitoring accuracy after forearm position changes, highlighting the danger of imprecise EMG-based neuromuscular monitoring during the administration of neuromuscular blocking agents. The study's strength lies in identifying Pattern C, where the ulnar nerve crosses the line connecting the anode and cathode, significantly enhancing monitoring accuracy. This novel attachment method holds promise to improve EMG-based neuromuscular monitoring precision in surgery involving forearm limb position changes, although further research is required to assess its utility comprehensively.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app