Comparison of a New Scheimpflug Camera and Swept-Source Optical Coherence Tomographer for Measurements of Anterior Segment Parameters.
Ophthalmology and Therapy 2023 September 26
INTRODUCTION: This study evaluated the differences and agreement between a new Scheimpflug camera (Scansys) and a swept-source anterior segment optical coherence tomographer (CASIA 2) for measurements of the anterior segment of the eye in normal subjects.
METHODS: This prospective study included 84 eyes from 84 normal adult subjects who underwent three consecutive measurements with the Scansys and the CASIA 2 in random order. The mean keratometry (Km), astigmatism magnitude (AST), J0 , and J45 vectors for both anterior and posterior corneal surfaces, central corneal thickness (CCT), thinnest corneal thickness (TCT), and anterior chamber depth (ACD) were obtained by both devices. The difference between these two devices was assessed using paired t test and violin plots. Bland-Altman plots and 95% limits of agreement (LoAs) were used to evaluate agreement.
RESULTS: No statistically significant differences between the two devices were found for the anterior AST, anterior J45 , and posterior J45 (P > 0.05). The remaining parameters were statistically significant (P ≤ 0.05), but the differences not clinically significant. The violin plots showed that the distribution and probability density of the measured parameters were similar for both devices. Bland-Altman plots revealed high agreement for the measured parameters between the Scansys and CASIA 2, with narrow 95% LoAs.
CONCLUSIONS: In terms of assessing parameters for the anterior segment, our study indicated that Scansys and CASIA 2 generally showed significant agreement. The two devices used in this study's assessment of all the parameters can be used interchangeably in refractive analysis.
METHODS: This prospective study included 84 eyes from 84 normal adult subjects who underwent three consecutive measurements with the Scansys and the CASIA 2 in random order. The mean keratometry (Km), astigmatism magnitude (AST), J0 , and J45 vectors for both anterior and posterior corneal surfaces, central corneal thickness (CCT), thinnest corneal thickness (TCT), and anterior chamber depth (ACD) were obtained by both devices. The difference between these two devices was assessed using paired t test and violin plots. Bland-Altman plots and 95% limits of agreement (LoAs) were used to evaluate agreement.
RESULTS: No statistically significant differences between the two devices were found for the anterior AST, anterior J45 , and posterior J45 (P > 0.05). The remaining parameters were statistically significant (P ≤ 0.05), but the differences not clinically significant. The violin plots showed that the distribution and probability density of the measured parameters were similar for both devices. Bland-Altman plots revealed high agreement for the measured parameters between the Scansys and CASIA 2, with narrow 95% LoAs.
CONCLUSIONS: In terms of assessing parameters for the anterior segment, our study indicated that Scansys and CASIA 2 generally showed significant agreement. The two devices used in this study's assessment of all the parameters can be used interchangeably in refractive analysis.
Full text links
Related Resources
Trending Papers
Renin-Angiotensin-Aldosterone System: From History to Practice of a Secular Topic.International Journal of Molecular Sciences 2024 April 5
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app