Integrated Time Nanosecond Pulse Irreversible Electroporation (INSPIRE): Assessment of Dose, Temperature, and Voltage on Experimental and Clinical Treatment Outcomes.

OBJECTIVE: This study sought to investigate a novel strategy using temperature-controlled delivery of nanosecond pulsed electric fields as an alternative to the 50-100 microsecond pulses used for irreversible electroporation.

METHODS: INSPIRE treatments were carried out at two temperatures in 3D tumor models using doses between 0.001s and 0.1s. The resulting treatment zones were quantified using viability staining and lethal electric field intensities were determined numerically. Computational modeling was then used to determine parameters necessary for INSPIRE treatments to achieve equivalent treatment zones to clinical electroporation treatments and evaluate the potential for these treatments to induce deleterious thermal damage.

RESULTS: Lethal thresholds between 1109 and 709V/cm were found for nominal 0.01s treatments with pulses between 350ns and 2000ns at physiological temperatures. Further increases in dose resulted in significant decreases in lethal thresholds. Given these experimental results, treatment zones comparable to clinical electroporation are possible by increasing the dose and voltage used with nanosecond duration pulses. Temperature-controlled simulations indicate minimal thermal cell death while achieving equivalent treatment volumes to clinical electroporation.

CONCLUSION: Nanosecond electrical pulses can achieve comparable outcomes to traditional electroporation provided sufficient electrical doses or voltages are applied. The use of temperature-controlled delivery may minimize thermal damage during treatment.

SIGNIFICANCE: Intense muscle stimulation and the need for cardiac gating have limited irreversible electroporation. Nanosecond pulses can alleviate these challenges, but traditionally have produced significantly smaller treatment zones. This study suggests that larger ablation volumes may be possible with the INSPIRE approach and that future in vivo studies are warranted.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app