Reflex strategy to ensure accurate total testosterone results from consumer initiated, self-collected capillary samples.
BACKGROUND: Self-collected capillary samples are convenient for direct access testing (DAT), but exogenous testosterone use may cause falsely elevated total testosterone (TT) results. We designed a quality assurance workflow to differentiate between accurate or erroneous supraphysiological TT concentrations.
METHODS: Clinical samples with TT > 1500 ng/dL were reflexed to luteinizing hormone (LH) and follicle stimulating hormone (FSH) and screened for exogenous testosterone use. Samples (n = 120) with normal TT were reflexed to LH/FSH as a control.
RESULTS: A total of 8572 TT samples were evaluated, of which 533 (6.2 %) had TT > 1500 ng/dL and were reflexed. Of these, 441 (82.7 %) had significantly decreased LH/FSH (<0.85/<0.7mIU/mL, respectively), 72 (13.5 %) had normal or borderline normal LH/FSH, and 20 (3.8 %) had insufficient plasma volume. In patients with TT > 1500 ng/dL, injectable exogenous testosterone use was most commonly accompanied by significantly decreased LH/FSH, while topical testosterone use was most commonly accompanied by detectable LH/FSH. Control samples were almost all (99.2 %) within or above the LH/FSH reference intervals. Unique patients ordered 351 TT tests where at least one TT result was > 1500 ng/dL. Based on TT and LH/FSH results, we hypothesized that patients were intermittently or consistently overusing exogenous testosterone, resolved elevated TT with recollection, or repeatedly contaminated their sample.
CONCLUSION: Self-collected capillary specimens are acceptable for TT testing. A quality assurance reflex to LH/FSH can determine the validity of supraphysiological TT results in a consumer initiated/DAT population.
METHODS: Clinical samples with TT > 1500 ng/dL were reflexed to luteinizing hormone (LH) and follicle stimulating hormone (FSH) and screened for exogenous testosterone use. Samples (n = 120) with normal TT were reflexed to LH/FSH as a control.
RESULTS: A total of 8572 TT samples were evaluated, of which 533 (6.2 %) had TT > 1500 ng/dL and were reflexed. Of these, 441 (82.7 %) had significantly decreased LH/FSH (<0.85/<0.7mIU/mL, respectively), 72 (13.5 %) had normal or borderline normal LH/FSH, and 20 (3.8 %) had insufficient plasma volume. In patients with TT > 1500 ng/dL, injectable exogenous testosterone use was most commonly accompanied by significantly decreased LH/FSH, while topical testosterone use was most commonly accompanied by detectable LH/FSH. Control samples were almost all (99.2 %) within or above the LH/FSH reference intervals. Unique patients ordered 351 TT tests where at least one TT result was > 1500 ng/dL. Based on TT and LH/FSH results, we hypothesized that patients were intermittently or consistently overusing exogenous testosterone, resolved elevated TT with recollection, or repeatedly contaminated their sample.
CONCLUSION: Self-collected capillary specimens are acceptable for TT testing. A quality assurance reflex to LH/FSH can determine the validity of supraphysiological TT results in a consumer initiated/DAT population.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app