Comparing Equil patch versus traditional catheter insulin pump in type 2 diabetes using continuous glucose monitoring metrics and profiles.
Journal of Diabetes 2024 April
AIMS: It is not clear whether there are differences in glycemic control between the Equil patch and the MMT-712 insulin pump. Our objective was to compare two types of insulin pumps in the treatment of type 2 diabetes mellitus (T2DM), using continuous glucose monitoring (CGM) metrics and profiles.
METHODS: This was a randomized case-crossover clinical trial. Participants were hospitalized and randomly allocated to two groups and underwent two types of insulin pump treatments (group A: Equil patch-Medtronic MMT-712 insulin pump; group B: Medtronic MMT-712-Equil patch insulin pump) separated by a 1-day washout period. Glycemic control was achieved after 7-8 days of insulin pump therapy. Each patient received CGM for 5 consecutive days (from day 1 to day 5). On day 3 of CGM performance, the Equil patch insulin pump treatment was switched to Medtronic MMT-712 insulin pump treatment at the same basal and bolus insulin doses or vice versa. CGM metrics and profiles including glycemic variability (GV), time in range (TIR, 3.9-10.0 mmol/L), time below range (TBR, <3.9 mmol/L), time above range (TAR, >10.0 mmol/L), and postprandial glucose excursions, as well as incidence of hypoglycemia.
RESULTS: Forty-six T2DM patients completed the study. There was no significant difference in parameters of daily GV and postprandial glucose excursions between the Equil patch insulin pump treatment and the Medtronic insulin pump treatment. Similarly, there was no between-treatment difference in TIR, TBR, and TAR, as well as the incidence of hypoglycemia.
CONCLUSION: The Equil patch insulin pump was similar to the traditional MMT-712 insulin pump in terms of glycemic control. Equil patch insulin pump is a reliable tool for glycemic management of diabetes mellitus.
METHODS: This was a randomized case-crossover clinical trial. Participants were hospitalized and randomly allocated to two groups and underwent two types of insulin pump treatments (group A: Equil patch-Medtronic MMT-712 insulin pump; group B: Medtronic MMT-712-Equil patch insulin pump) separated by a 1-day washout period. Glycemic control was achieved after 7-8 days of insulin pump therapy. Each patient received CGM for 5 consecutive days (from day 1 to day 5). On day 3 of CGM performance, the Equil patch insulin pump treatment was switched to Medtronic MMT-712 insulin pump treatment at the same basal and bolus insulin doses or vice versa. CGM metrics and profiles including glycemic variability (GV), time in range (TIR, 3.9-10.0 mmol/L), time below range (TBR, <3.9 mmol/L), time above range (TAR, >10.0 mmol/L), and postprandial glucose excursions, as well as incidence of hypoglycemia.
RESULTS: Forty-six T2DM patients completed the study. There was no significant difference in parameters of daily GV and postprandial glucose excursions between the Equil patch insulin pump treatment and the Medtronic insulin pump treatment. Similarly, there was no between-treatment difference in TIR, TBR, and TAR, as well as the incidence of hypoglycemia.
CONCLUSION: The Equil patch insulin pump was similar to the traditional MMT-712 insulin pump in terms of glycemic control. Equil patch insulin pump is a reliable tool for glycemic management of diabetes mellitus.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app