Assessment of Microgap and Bacterial Leakage of Two Types of Internal Implant-Abutment Union.
Journal of Pharmacy & Bioallied Sciences 2024 Februrary
OBJECTIVE: The current research was done to assess microbial seepage of two types of internal implant-abutment connections.
MATERIALS AND METHODS: Twenty dental implants are categorized into two groups. Group A fixtures with an internal hexagonal geometry and group B fixtures with a tri-lobe internal connection. All implant-abutment assemblies underwent a three-week incubation period at 37°C in sterile tubes containing 5 mL of Staphylococcus aureus broth culture. Through the use of Gram stain and biochemical processes, the resultant colonies were recognized.
RESULTS: The mean Log10 colony forming unit (CFU) in group A was 8.4 and in group B was 7.2. The variation between both groups was found to be considerable ( P < 0.05). Microgap was more in group B compared to group A.
CONCLUSION: Bacteria may infiltrate the small area between the implant and the abutment. Compared to dental implant fixtures with a tri-lobe internal connection, there was a noticeably higher Log10 CFU in dental implant fixtures with an internal hexagonal geometry.
MATERIALS AND METHODS: Twenty dental implants are categorized into two groups. Group A fixtures with an internal hexagonal geometry and group B fixtures with a tri-lobe internal connection. All implant-abutment assemblies underwent a three-week incubation period at 37°C in sterile tubes containing 5 mL of Staphylococcus aureus broth culture. Through the use of Gram stain and biochemical processes, the resultant colonies were recognized.
RESULTS: The mean Log10 colony forming unit (CFU) in group A was 8.4 and in group B was 7.2. The variation between both groups was found to be considerable ( P < 0.05). Microgap was more in group B compared to group A.
CONCLUSION: Bacteria may infiltrate the small area between the implant and the abutment. Compared to dental implant fixtures with a tri-lobe internal connection, there was a noticeably higher Log10 CFU in dental implant fixtures with an internal hexagonal geometry.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app